- 23 -
PEAK
QQ2
: the peak function is run only via input (default);
<(6
: the peak function is also run via the START/STOP key.
By keeping the PEAK input closed or pressing the START/STOP key (if enabled), the maximum
weight value reached remains displayed. By opening the input or keeping the START/STOP key (if
enabled) pressed for 3 seconds, the current weight is displayed.
If you wish to use this input to view a sudden variation peak, set the FILTER ON THE
WEIGHT to 0.
ANALOG OUTPUT(ONLY FOR INSTRUMENTS WHERE THIS OPTION IS AVAILABLE)
-
7<3(
: it selects the analog output type (4÷20 mA, 0÷20 mA, 0÷10 V, 0÷5 V,
±
10 V,
±
5 V;
default: 4÷20 mA).
For the output
±
10 V and
±
5 V the soldered jumper SW1 must be closed:
▫
open the instrument, releasing with a screwdriver the locking tabs that hold together
the two sides of the case;
▫
locate on the printed circuit board the soldered jumper SW1 highlighted in the picture
below:
▫
close the jumper shorting the pads with a drop of tin.
-
02'(
: choice of a weight followed by the analog output: gross (
*5266
) or net (
1(7
). If the net
function is not active, the analog output varies according to gross weight.
-
$1$
: set the weight value for which you wish to obtain the minimum analog output value.
Only set a value different from zero if you wish to limit the analog output range; for
instance: for a full scale value of 10000 kg you require an 4 mA signal at 5000 kg and
20 mA at 10000 kg, in this case, instead of zero, set 5000 kg.
+
&$/,E
3($.21
+
&$/,E
$Q$/2*