- 3 -
LOAD CELL INPUT TEST (QUICK ACCESS)
From the weight display, press
for 3 seconds: the top of the
display shows the gross weight; the bottom shows the response
signal of each load cell expressed in mV with three decimals.
Example: a load cell with 2.000 mV/V sensitivity provides a
response signal between 0 and 10 mV.
LOAD CELL TESTING
Load cell resistance measurement (use a digital multimeter):
-
Turn off the instrument.
-
The value between the positive signal wire and the negative signal wire must be equal or similar
to the one indicated in the load cell data sheet (output resistance).
-
The value between the positive excitation wire and the negative excitation wire must be equal or
similar to the one indicated in the load cell data sheet (input resistance).
Load cell voltage measurement (use a digital multimeter):
-
Turn on the instrument.
-
Take out the load cell to be tested from underneath the container, or alternatively, lift the
container support.
-
Make sure that the excitation of two wires of the load cell connected to the instrument (or
amplifier) is 5 VDC ±3%.
-
Measure the response signal between the positive and the negative signal wires by directly
connecting them to the tester, and make sure that it is comprised between 0 and 0.5 mV.
-
Apply load to the cell and make sure that there is a signal increment.
IF ONE OF THE ABOVE CONDITIONS IS NOT MET, PLEASE CONTACT THE TECHNICAL
ASSISTANCE SERVICE.
3 s