- 15 -
CONFIRMATION AND CHANGE OF ACTIVE CHANNELS
After performing the calibration and verifying that the system works properly, you can confirm the
channels automatically detected by the instrument; in this way, in case of accidental interruption of
the cable of one or more load cells, the instrument displays the
alarm.
Automatic load cells detection is enabled by default on all 8 channels of the instrument.
: the display shows the automatically detected channels;
press ESC to exit with no change; press ENTER to make the
current selection permanent and disable automatic detection.
In case a channel is not working, you can move the load cell on a
free channel: in this screen press
or
to select the new
channel, press ENTER and select with
o
or
to activate or deactivate it.
After editing, you must repeat equalization, zero setting and calibration using a sample
weight.
: manual selection of active channels is cancelled; automatic detection on all channels is
enabled. Confirmation is requested (
),press ENTER to proceed or press ESC to cancel.
+
+