Proline Low-temperature Thermostats
YACE0072 / 21.08.07
Starting up
41
7.6.6 Submenu
LMDkt
Æ
O0c--
(Master): Setting external control parameters
You only obtain the following possible settings when the control variable is set to "External" (
⇒
Section
*Display flashes. Set
with or .
P-contr. range: 3K … 250K.
P-controller [K]
Da 2/
°C
*Display flashes. Set
with or .
Tve range: Off, 1s … 800s.
Deriv time master cntr [s]
Ds013
°C
Display flashes. Set
with or .
Tne range: 5s … 979s, Off.
Res time master cntr [s]
Dl04/
°C
Set control parameters
O0c--
°C
Display flashes. Set
with or .
Kpe range: 0,10 … 9.80.
Prop.rng.master contr.[ ]
DO<4/
°C
* Important:
The windows for setting the
additional parameters for the
external control for the
Master
Controller:
Ds
(=tve) and
Dc
(=tde),
and for the
Slave Controller
:
Da
(=P-controlller),
are only displayed when
D
has
been changed from the works
setting of "Automatic" to
"Manual".
Changeover should only be
carried out by experienced
control technicians.
*Display flashes. Set
with or .
Tde range: 0.0s … 50.0s.
Damping master cntr [s]
Dc0!3
°C
Display flashes. Set
with or .
Xpf range: 0.3K … 25K.
Prop. range slave contr[K]
hO $/
°C
Continued on the
following page
Summary of Contents for RP 845
Page 2: ......