Proline Low-temperature Thermostats
YACE0072 / 21.08.07
Starting up
69
−
The temperature measurement device
shows the true temperature value (with
glass thermometers consider the
correction where applicable!).
−
Change the display in the adjacent
window to the true value with cursor or
soft keys and accept with
or End ,
−
or quit the window with
without making
changes.
7.8.8 Restoring the works setting of the internal temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Master
cDE 0
3x
to
LMDls
2x
to
B@K--
1x
−
Call offset works setting for the internal temperature probe
cDE
0
.
−
cDE 0
is displayed.
3 sec
−
The works setting is accepted after pressing for 3 seconds and
cmlD
is displayed.
Command
Default
−
Open the device parameter menu via
the soft key Menu .
−
With the cursor keys continue to
Æ
Settings
Æ
Calibration
Æ
intern Pt100
Æ
Default
.
−
The window shown adjacent appears.
−
Confirm selection with
.
cDE 0
°C
Help Menu
Screen T
set
T
fix
6 7 8 9
0
1 2 3 4
5
Temperature value of the ref.
temp. measurement device:
20.
15
Min: -50.0°C Max: 302.00
Help Menu
End T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default