Lauda PD 2 G Operating Instructions Manual Download Page 17

 

09/2020 

Puridest Glass Distillation Apparatus 

17 / 34 

 
 

 

PD 8 G 

 

Exterior dimensions (W x D x H) 

685 mm x 360 mm x 410 mm 

 

Distillation capacity 

8 l / h mono distillate 

 

Distillate quality 

Mono distillate 
approx. 2.2 µS / cm at 25 °C 
in conformity with DAB, bacteria and pyrogen free, 
low gas content. 

 

The conductivity of the distilled water is directly related to the chemical composition of the raw water. Components of the raw water having the same 
or a lower evaporation point than water may deteriorate the conductivity. 

 

 

 

Cooling water required 

144 l / h 

 

Low water cut-off 

Electromechanical temperature limiter, with capillary 
tube sensor. 
Switch-off temperature 135 °C / -15 K 

 

Water pressure min. / max. 

> 3 bar / 7 bar 
> 43.5 psi / 101.5 psi 

 

Electrical connection 
Mains supply 
Internal fuse 
On-site fuse 

220 V / 3 ~ / PE, +/-10 %, 50 / 60 Hz, 6.0 kW 
Mains connection cable for permanent connection to 
the mains, on-site 3 x 25 A inert 
or 
400 V / 3 ~ / N / PE, +/-10 %, 50 / 60 Hz, 6.0 kW 
Mains connection cable for permanent connection to 
the mains, on-site 3 x 16 A inert 

 

Caution! Mains voltage deviations, even with the mentioned limits, influence the quantity of distillate produced. 

Protection class / - type 

I / IP20 

 

 

 

 

Surrounding conditions 

Use only inside buildings (not in potentially explosive 
areas). 

 

Height above MSL 

up to 2000 m MSL 

 

Ambient temperature 

+10 °C to +40 °C 

 

Humidity 

max. 80 % rel. humidity to 31 °C, decreasing to 40 % 
rel. humidity at 40 °C 

 

 

 

 

Net Weight 

26.0 kg 

 

Weight with water filling 

34.0 kg 

 

 
 
 

 

Summary of Contents for PD 2 G

Page 1: ...Operating Instructions Puridest Glass Distillation Apparatus PD 2 G PD 4 G PD 8 G PD 2 DG PD 4 DG...

Page 2: ...2 34 Puridest Glass Distillation Apparatus 09 2020...

Page 3: ...i distillate The distillate is in conformity with DAB regulations and the regulations of international pharmacopoeia Before installation please check whether contents of package are in good order and...

Page 4: ...4 34 Puridest Glass Distillation Apparatus 09 2020...

Page 5: ...tional Description 12 Puridest Glass Distillation Apparatus 12 Protection against over heating in case of low water 12 Glass Distillation Apparatus with separate water supply 12 Glass Distillation App...

Page 6: ...the mains supply examples 25 13 1 1 PD 2 G PD 4 G and PD 2 DG for 230 V at Mains supply 220 V 3 PE 50 60 Hz 25 13 1 2 PD 2 G PD 4 G and PD 2 DG for 230 V at Mains supply 220 V 3 PE 50 60 Hz 25 13 1 3...

Page 7: ...rox 1 6 S cm at 25 C The quality of the distilled water is among other things dependent on the composition of tap water as well as on the cleanliness of the distillation stages The distillate quality...

Page 8: ...persons The Distillation Apparatus can be held between the four stands of the unit and lifted for setup The unit is suitable for both bench and wall mounting It is only suitable for indoor use Table...

Page 9: ...e required mains fuse see Technical Data of these instructions The electrical connection must ensure an all pole separation from the mains at any time Shock proof sockets or main switches to separate...

Page 10: ...omestic water connection for pretreated water with a pressure hose with lockable water connection cut off valve Both hose connections must by all means be secured with hose clips Tap water inlet The t...

Page 11: ...ratures up to 100 C Danger of scaldings Waste water drain When the function Clean water exchange is activated impure water drains from the boiler through the connection waste water drain Connect a hos...

Page 12: ...te can be withdrawn from the distillate outlet of Bi Glass Distillation Apparatus Protection against over heating in case of low water Float switches control the water level in the boilers Power to th...

Page 13: ...a temperature of maximum 70 C switch the Distillation Apparatus off Make absolutely sure not to bring the solution to a boil Allow to react for approx 1 to 2 hours with the scale deposits Af ter the r...

Page 14: ...n is presumably always due to a defect of the unit Models PD 2 G and PD 4 G have only one low water cut off Pilot lamp Clean Depending on the degree of impurities in the tap and the increasing contami...

Page 15: ...sks by the person in charge company date signature 10 Disposal of Old Units LAUDA GFL will take responsibility within the scope of the legal directives for an environmentally sound handling and dispos...

Page 16: ...emperature 135 C 15 K Electromechanical temperature limiter with ca pillary tube sensor Switch off temperature 135 C 15 K Water pressure min max 3 bar 7 bar 43 5 psi 101 5 psi 3 bar 7 bar 43 5 psi 101...

Page 17: ...e sensor Switch off temperature 135 C 15 K Water pressure min max 3 bar 7 bar 43 5 psi 101 5 psi Electrical connection Mains supply Internal fuse On site fuse 220 V 3 PE 10 50 60 Hz 6 0 kW Mains conne...

Page 18: ...illary tube sensor Switch off temperature 135 C 15 K Water pressure min max 3 bar 7 bar 43 5 psi 101 5 psi 3 bar 7 bar 43 5 psi 101 5 psi Electrical connection Mains supply On site fuse 230 V 10 50 60...

Page 19: ...G F3 Low water cut off front boiler stage F4 Low water cut off rear boiler stage F5 Fuse 400 mA inert on A1 H1 Pilot lamp Cleaning H2 Pilot lamp Sterilising H3 Pilot lamp Low Water K1 Contactor mains...

Page 20: ...20 34 Puridest Glass Distillation Apparatus 09 2020 PD 2 G for 230 V 50 60 Hz PD 4 G for 230 V 50 60 Hz...

Page 21: ...09 2020 Puridest Glass Distillation Apparatus 21 34 PD 8 G for 220 V 3 PE 50 60 Hz PD 8 G for 400 V 3 N PE 50 60 Hz...

Page 22: ...22 34 Puridest Glass Distillation Apparatus 09 2020 PD 2 DG for 230 V 50 60 Hz PD 4 DG for 220 V 3 PE 50 60 Hz...

Page 23: ...09 2020 Puridest Glass Distillation Apparatus 23 34 PD 4 DG for 400 V 3 N PE 50 60 Hz Circuit diagram for accessory level control switch and option separate water supply...

Page 24: ...N PE 50 60 Hz ge gr yellow green PE Protective earth PE Protective earth bl blue N sw black L1 L1 br brown L2 L2 gra grey L3 L3 Glass Distillation Apparatus can be supplied in different versions for...

Page 25: ...on site S4 Main switch on site 13 1 1 PD 2 G PD 4 G and PD 2 DG for 230 V at Mains supply 220 V 3 PE 50 60 Hz see nameplate PD 2 G PD 4 G and PD 2 DG for 230 V Mains connection through CEE plug 7 7 pr...

Page 26: ...ameplate PD 8 G and PD 4 DG Permanent mains connection through on site switch S4 all pole sepa ration from the mains possible 13 1 4 PD 8 G and PD 4 DG for 220 V 3 at Mains supply 400 V 3 N PE 50 60Hz...

Page 27: ...DG Permanent mains connection through CEE plug according to IEC stand ard 60309 2 all pole separation from the mains possible 13 1 6 PD 8 G and PD 4 DG for 220 V 3 at Mains supply 400 V 3 N PE 50 60 H...

Page 28: ...28 34 Puridest Glass Distillation Apparatus 09 2020 14 List of spare parts When ordering spare parts please always state type and serial number of the unit in question...

Page 29: ...0024120 Seat for float switch Option separate water supply 12 0024115 Siphon with overflow 13 A000868 Shaft tube 14 A000936 Overflow 15 0024114 Glass angle 16 0024122 Boiler models PD 2 G PD 2 DG A00...

Page 30: ...del of the Distillation Apparatus Separate Water Supply with solenoid valve Only factory installation into all Glass Distillation Apparatus is possible When the storage tank is full the solenoid valve...

Page 31: ...s for water inlet and outlet length 1 5 m including hose clips Hose Set Hose Set for models PD 2G to PD 8 G for models PD 2 G to PD 8 G for seperate water supply Part No A000139 Part No A000143 Hose S...

Page 32: ...32 34 Puridest Glass Distillation Apparatus 09 2020...

Page 33: ...09 2020 Puridest Glass Distillation Apparatus 33 34...

Page 34: ...LAUDA GFL Gesellschaft f r Labortechnik mbH Schulze Delitzsch Stra e 4 5 30938 Burgwedel Germany Tel 49 0 5139 9958 0 Fax 49 0 5139 9958 21 E Mail info lauda gfl de Internet www GFL de...

Reviews: