Designation
Container size
Catalog number
Kryo 51
LZB 121
LZB 221
LZB 321
Kryo 30
LZB 109
LZB 209
LZB 309
Kryo 20
LZB 116
LZB 216
LZB 316
Aqua 90
LZB 120
LZB 220
LZB 320
Therm 250
LZB 122
LZB 222
LZB 322
Therm 180
LZB 114
LZB 214
LZB 314
Therm 160
LZB 106
LZB 206
LZB 306
When using Kryo 30:
The water content decreases during longer periods of operation at
higher temperatures and the mixture becomes combustible (flash point
119°C). Check the mixing ratio using a hydrometer.
When using Aqua 90:
Evaporation losses occur at higher temperatures. In this case, use a bath
cover.
Never use silicone oil in silicone tubes.
When using mineral oils:
Do not use in combination with an EPDM hose.
The proportion of alkaline earth ions in the water must be between
0.71 mmol/L and 1.42 mmol/L (equivalent of 4.0 and 8.0 °dH). Harder
water leaves limescale deposits in the device.
The pH value of the water must be between 6.0 and 8.5.
Distilled, deionized and demineralized water are unsuitable due to their
reactivity. High-purity water and distillates are suitable as a heat transfer
liquid after 0.1 g of soda (Na
2
CO
3
, sodium carbonate) is added for every
liter of water.
Sea water is unsuitable due to its corrosive properties.
Avoid chlorine content in the water at all costs. Do not add chlorine
to the water. Cleaning agents and disinfectants, for example, contain
chlorine.
The water must be free of impurities. Water with iron content is unsuit-
able due to rust formation and untreated river water is unsuitable due to
algae growth.
The addition of ammonia is not permitted.
5.2 Filling the device
LAUDA is not liable for damages resulting from the use of unsuitable heat
transfer liquids. Approved heat transfer liquids
Heat transfer liquid, water
V6
PRO bath thermostats and circulation thermostats
60 / 156
Summary of Contents for P 10
Page 146: ...V6 PRO bath thermostats and circulation thermostats 146 156 ...
Page 147: ...V6 PRO bath thermostats and circulation thermostats 147 156 ...
Page 148: ...V6 PRO bath thermostats and circulation thermostats 148 156 ...
Page 157: ......
Page 158: ......
Page 159: ......