Fig. 21: Entries for the test
7.
A command must be sent to the constant temperature equipment
to test communication. For example,
TYPE
. Type the command and
press [Send ASCII].
If the connection is operational, the constant temperature equip-
ment acknowledges the command.
4.4.3 Data transfer rate
The data transfer rate cannot be precisely defined. It depends on a variety of
different factors:
Are the constant temperature equipment ( with the Ethernet interface)
and the control station/PC on the same network?
Are the control station/PC and constant temperature equipment con-
nected wirelessly (WiFi) or through a cable?
What is the load on the network?
Generally commands can be sent to the constant temperature equipment
every 500 ms. For WiFi connections the period could exceed 1 s. A new
command can only be sent if the previous command has been acknowledged
by the constant temperature equipment.
V6
PRO bath thermostats and circulation thermostats
43 / 156