background image

OPERATION MANUAL

4 channel  

thermometer with SD logger and relay

88597(4 channel thermistor )

88599(2 channel K, 2 channel thermistor)

www.ete.co.th

Summary of Contents for 88597

Page 1: ...OPERATION MANUAL 4 channel thermometer with SD logger and relay 88597 4 channel thermistor 88599 2 channel K 2 channel thermistor www ete co th ...

Page 2: ...e Hold function freezes current readings Checking MAX MIN value Beeper and LED alarm are both available One press to display T1 T2 Display temperature change from baseline Programmable temperature high low alarm threshold Temperature offset function for adjustment purpose Over range indication with error messages Blue backlight for dark area operation 12hour 24hour time format selectable Real time...

Page 3: ... operation and reading When battery level is lower than certain range the logging function will be terminated If the low battery icon appeared on the LCD please remove the battery cover from the rear side and replaced 4 new AA batteries into the battery compartment Please check the polarity and make good contact Caution 1 The battery life will be shorten when using a high capacity SD card or in hi...

Page 4: ...e reading was adjusted manually 7 The MEMO icon indicates Manual logging is in processing 8 The REC icon indicates Auto logging is in processing 9 The RES icon indicates the logger is in reservation status 10 The CH1 CH2 CH3 CH4 display probe temperature 11 The Hi Lo icon indicate the high or low threshold is over 12 The AM PM icon indicate the time format 13 The SD icon indicates the SD card is i...

Page 5: ...nsert the SD card by following the icon reminder on housing To take out the SD card press the SD card to eject 4 Adaptor socket Bottom View REAR View 5 Relay port 6 Probe socket For 88599 2 Miniature K probe sockets 2 thermistor probe sockets For 88597 4 thermistor probe sockets 1 1 Red LED alarm indicator Keypad 5 ...

Page 6: ...ince power on 1 Press key to see maximum and minimum value of each channel in turns 2 The MAX MIN value will be reset after power off 3 The MAX MIN will not be recorded if the ERROR message displayed 4 This mode is also working in above REL status HOLD Press key to freeze the readings Press again to release and return to normal measurement mode T1 T2 To view the temperature difference between CH1 ...

Page 7: ...he meter is REC status auto power off function is disable 7 When the meter is in MAX MIN HOLD T1 T2 REL status the function will be dismissed after pressing key 8 When the meter is in recording status press power key can not turn the meter off The recording mode must be stopped first to turn the meter off AUTO LOGGING make reservation to start 1 While the logger is pre programmed as reservation st...

Page 8: ...m issue happened already To eliminate the Lo or Hi icon from LCD press 2 seconds MANUAL LOGGING record for one data 1 Pressing key the MEMO symbol will be appeared on the LCD to indicate the manual recording is in process 2 It will record 1 data and return to the measurement mode 3 The shortest interval for taking manual memory is 5 seconds 4 Data recording is unlimited 5 If the meter is in either...

Page 9: ...TERY ICON 1 When the battery is low the low battery icon will appear on the display Suggest to change the batteries 2 The Manual logging or Auto logging is invalid when the low battery icon appears on the display 3 When the low battery icon appears on the LCD the reservation function will not function 4 If the low battery icon appears during the recording mode the data will be recorded until the p...

Page 10: ...ore than 2 sec will return to normal measurement mode 8 The details of each setting up function is listed in next page SET UP The advanced SETUP function lets you customize your meter s preferences and defaults The programmable parameters are 1 RECORDING SETTING Auto logging related setup such as reservation start and sampling rate 1sec 2 sec 5sec 10sec 15sec 30sec 1min 2min 5min 10min 15min 30min...

Page 11: ...choose the channel and decide high or low alarm threshold Then each high or low alarm can be enabled or disable by choosing ON or OFF The previous alarm threshold value is stored in meter the last step is to setup the threshold The adjustable range is 200C to 1370C First choose the channel Then each set value can be enabled or disable by choosing ON or OFF The previous offset value is stored in me...

Page 12: ...s under 10M byte SD icon flashes You can t reserve setting or start recording if SD icon is flashing FULL icon appears when the SD card capacity is full SD CARD PREPARATION This product is compatible format SD card Suggest SD card capacity is 8GB SD SDHC card or below Before loading or unloading the SD card into meter please follow the suggestion below 1 The data loss caused by the damaged SD card...

Page 13: ...s from Jan 25th 13 24pm the file name is 01251324 txt The records will be covered if the recording time is same as one year ago 2 MANUAL LOGGING File time Power on made the file The data will record on the same file File name azxxx tet such as power on Jan 26 13 24pm the file name is 01261324 txt The recorded data is stored in SD card in the following format The temperature unit is determined by s...

Page 14: ...10 Solution Put the meter in regular room for 30 minutes If the error message still appear Send back for repair E08 The room Temp is higher than 60 Solution Put the meter in regular room for 30 minutes If the error message still appear Send back for repair E09 This error message indicates he reservation start time or stop time setting is wrong Solution review setting E10 SD card does t insert or S...

Page 15: ...be 200 1370 328 2498 0 1 0 1 0 3 rdg 1 Programmable from 1 second up 47x104 0 50 Humidity 80 20 50 Humidity 90 152x100x39 300g 4pcs AA alkaline battries or 9V adaptor Meter 2pcs bead type K thermocouple 2 or 4 thermistor probes battery manual SD card plain box K type thermocouple custom made thermocouple is available upon request Thermistor temp range Thermistor resolution Thermistor accuracy 0 1 ...

Page 16: ...s Warranty is void if the meter has been opened Authorization must be obtained from the supplier before returning items for any reason When requiring a RA Return Authorization please include data regarding the defective reason the meters are returned along with good packing to prevent any damage in shipment and insured against possible damage or loss RETURN AUTHORIZATION www ete co th Cape Instrum...

Page 17: ...nemometer Sound Level Meter Air Flow meter Infrared Thermometer K type Thermometer K J T type Thermometer K J T R S E type Thermometer pH Meter Conductivity Meter T D S Meter D O Meter Saccharimeter Manometer Tacho Meter Lux Light Meter Moisture Meter Data logger Temp RH transmitter Wireless Transmitter 2018 12 V01 ...