9 Faults
If a fault occurs, the interface distinguishes between different message types,
e.g. alarms, errors and warnings. The procedure for rectifying a fault depends
on the device. Follow the corresponding instructions in the operating manual
accompanying the constant temperature equipment.
If you are unable to rectify a fault, please contact LAUDA Service,
Chapter 1.6 “Contact LAUDA” on page 6
.
9.1
Error
The Pt100/LiBus modules recognize the following error messages:
Code *
Meaning
1701 – 1704 /
2001 – 2004
Interface module hardware faulty. Contact the LAUDA department.
1705 / 2005
Clock frequency incorrect.
1706 / 2006
Distribution voltage too low.
1707 / 2007
Distribution voltage too low.
1708 / 2008
Backup memory faulty.
1710 / 2010
2-point calibration: Upper / lower point mixed up.
1711 / 2011
2-point calibration: Difference between upper / lower measuring point too small.
1712 / 2012
2-point calibration: Difference between upper / lower correction point too small.
1713 / 2013
2-point calibration: Difference between upper measuring / correction point too great.
1714 / 2014
2-point calibration: Difference between lower measuring / correction point too great.
1724 / 2024
Product line detection faulty.
* Error prefix 17 applies for interface Pt100, error prefix 20 for Pt100-2, see
Chapter 4.3 “Coding switch” on page 11.
V1
Interface module LRZ 918 / 925
19 / 25