Table 6
Unit
IN 2050 PW
IN 2560 PW
Filling volume
‑
minimum
L
11.1
12.1
‑
Maximum
L
36.3
48.1
Pump data 50/60 Hz
‑
maximum discharge pressure
bar
6.0
6.0
‑
Maximum flow rate
l/min
120.0
100.0
Connector thread (outer) inlet/outlet
mm
M38 x 1.5
M38 x 1.5
Noise level 50 Hz
Â
dB(A)
58
74
Noise level 60 Hz
Â
dB(A)
58
74
Weight
kg
382.0
647.0
Distance between device and environment
‑
Front
mm
200
200
‑
Back
mm
200
200
‑
Right
mm
200
200
‑
Left
mm
200
200
Class division according to DIN 12 876-1
‑
Class designation
---
I
I
‑
Identification code
---
NFL (suitable for
non-flammable liquids)
NFL (suitable for
non-flammable liquids)
À
- ACC area (Active Cooling Control) according to DIN 12876 is the
working temperature range during operation with an active cooling unit.
Á
- Temperature stability determined according to standard DIN 12876-2
Â
- Noise level determined according to standard EN 11201 for operating
position in front of the device at 1 meter distance
15.2
Heating output and power supply
If the maximum current consumption is limited
consumption” on page 101 the heating output can be reduced.
Depending on the country-specific version, the maximum current con-
sumption may already be limited ex works. Limited current consumption
values are shown in brackets.
Table 1
Unit
IN 130 T
IN 230 T
IN 230 TW
Heating output
230 V; 50 Hz
kW
2.7
2.7
2.7
200 V; 50/60 Hz
kW
2.2
2.2
2.2
Integral T
V6
Integral Process Thermostats and High-Temperature Thermostats
166 / 198