2.
Select one of the following options:
n
With [Off], the device is set to standby operating mode after a
power failure and when the power supply is restored.
n
With [On], the device is set automatically continue after a power
failure and when the power supply is restored.
Automatic running of the device may result in unattended
operation.
3.
Press the Enter key to confirm your selection.
The new setting is active.
The fuse on the installation side must at least correspond to the maximum
current consumption of the device (see rating label). If the mains fuse is
lower, you must reduce the maximum current consumption of the device.
The heating output will be reduced accordingly. When setting the current
consumption, consider whether other loads may be connected to a fuse
together with your device.
Personnel:
n
Specialized personnel
1.
In the [Basic setup] menu select the menu item
Curr. Limit
.
An entry window opens for a numerical value. The cursor flashes
under the numerical value. The upper and lower entry range is
displayed.
2.
Change the value with the arrow keys.
By pressing the
[ESC]
key, you return to the
[Basic setup]
menu without making any changes.
3.
Press the [OK] to confirm the new value.
The new value is active.
The reduction of the maximum current consumption of the device
reduces the heating power and thus influences the control charac-
teristics, where applicable.
The menu languages English, German, French, Spanish, and Italian are avail-
able for the device display.
Limiting the current consumption
Selecting the menu language
V6
Integral Process Thermostats and High-Temperature Thermostats
101 / 198