Starting up
10.08.99
YACE0060
25
6.4.4.2 Indicating/ altering of programme segments
−
When PGM appears on the display (therefore proceed as
described in 6.4.4.) press the key on the left.
−
The display shows RUN. Here the programmer can be
started (
➾
Section 6.4.4.4.)
−
Forward with key until EDIT appears..
−
Here the programme segments can e.g. be indicated and
altered. To do so press the key.
−
The display indicates 3 variables: segment number (a),
segment end temperature (b) and segment time (c).
−
EXAMPLE
:
segment number 1, b = 30,0 °C, c = 20 min.
The bath liquid has to be heated up or cooled down to
30 °C within 20 min.
☞
−
When having deleted the programme example the variables on the display show 0.
Before altering the segments it is necessary to insert new segments
(
➾
Section 6.4.5.5.)
or
−
These keys can be used to scroll through the different
variables.
−
Sequence with the key: : 1 (a)
!
30,0 °C (b)
!
20 (c)
!
50,0 ° (2. Segment),
!
20 (2. Segment).
−
When the segment number is flashing, pressing the key
leads directly to the menu for inserting or deleting
segments (
➾
Section 6.4.4.5.). Otherwise scroll e.g. from
segment time back to segment temperature and segment
number.