Unit description
10.08.99
YACE0060
9
Permitted data formats
-XXX.XX
-XXX.X
-XXX.
-XXX
XXX.XX
XXX.X
XXX.
XXX
-XX.XX
-XX.X
-XX.
-XX
XX.XX
XX.X
XX.
XX
-X.XX
-X.X
-X.
-X
X.XX
X.X
X.
X
-.XX
-.X
.XX
.X
3.7.4
Data request from the thermostat
The response from the thermostat is always in the fixed format "XXX.XX" for negative
values "-XXX.XX" or " ERR_X" (RS 485 interface: ”A015_XXX.XX”, for negative values ”A015_-XXX.XX”
or ”A015_ERR_X”).
Input command
Explanation
IN_PV_00
Request to indicate the bath temperature
IN_SP_00
Request to indicate the active setpoint
IN_SP_01
Request to indicate the pump output step
IN_SP_02
Request to indicate the operating mode of the refrigeration
system “0“ = OFF, "1" = ON, “2“ = ON 50 % (reduced capacity
RE 207, 212, 220, 307, 312, 320 only) , "3" = Automatic
operation
IN_SP_03
Request to indicate the current overtemperature switch-off
point
IN_PAR_00
Request to indicate the current value of Xp
IN_PAR_01
Request to indicate the current value of Tn
TYPE
Request to indicate the unit type
VERSION
Request to indicate the software version number
STATUS
Request to indicate the unit status
0 = OK, - 1 = fault
STAT
Input for fault diagnosis
answer XXXXX
!
X = 0 no fault, X = 1 fault
char. 1 from right = internal fault microcontroller 2
char. 2 from right = internal fault microcontroller 1
char. 3 from right = pump blocked
char. 4 from right = low level
char. 5 from right = overtemperature
Summary of Contents for Ecoline RE 320
Page 2: ......
Page 8: ...Brief operating instructions 2...
Page 59: ...Circuit diagram 10 08 99 YACE0060 53 11 Circuit diagram...
Page 60: ...Circuit diagram 54...
Page 63: ...Pipe plan 16 08 99 YACE0060 57 12 Pipe plan...
Page 64: ...Pipe plan 58...
Page 65: ...Pipe plan 16 08 99 YACE0060 59...
Page 68: ......