LaserPoint s.r.l.
–
Cronos Instruction Manual
–
All Rights are Reserved
6
Note: to increase the battery life, the instrument automatically switches off after
5 minutes of stand by operation. However the last recorded value is not lost; to
retrieve it, go back to Paragraph 2 and follow the instructions. After the internal
auto check cycle, the display will show the last measured value.
To intentionally shut the instrument off, keep the
ON/Zero
button pushed until
the LCD shuts off (this takes approximately 4 seconds).
9-Out of Range Notice
When the head is exposed to a source exceeding its
full scale, the LCD shows “
ERR
” instead of a
numeric value
.
10-Wavelength Selection
To select the wavelength of operation (from CO2 to
YAG or viceversa), go back to Paragraph 2 (shut the
instrument off), but keep the
ON/Zero
button pushed
until the wavelength change appears on the display
(this step takes approximately 7 seconds). At this
point release the button and the instrument is ready for a new measurement.. The
displayed value is the last recorded power measurement carried out at this
wavelength setting.
11-Customer’s calibration
This instrument can be re-calibrated by the user to
match a custom reference. Remove the plastic cover
of the adjustment switch located at the right side of
the instrument body. Zero the instrument as
indicated in Paragraph 2 of this Operation Section
and afterwards carry out a measurement of a known laser source. Shift the switch
on + or – if you need to increase or decrease the reading displayed on the LCD to
match your reference. This is done by limited pushes on the
ON/Zero
key, until
the desired power value is achieved on the display. Reset the switch to its middle
position
CAL
and re-cap. The instrument is ready for new measurements.