Cronos Instruction Manual

 

 All Rights are Reserved

  

Note: to increase the battery life, the instrument automatically switches off after 
5 minutes of stand by operation. However the last recorded value is not lost; to 
retrieve it, go back to Paragraph 2 and follow the instructions. After the internal 
auto check cycle, the display will show the last measured value. 
To intentionally shut the instrument off, keep the 

ON/Zero

 button pushed until 

the LCD shuts off (this takes approximately 4 seconds). 
 

9-Out of Range Notice 

When  the  head  is  exposed  to  a  source  exceeding  its 
full  scale,  the  LCD  shows  “

ERR

”  instead  of  a 

numeric value

.

 

10-Wavelength Selection

 

To select the wavelength of operation (from CO2 to 
YAG or viceversa),  go back to Paragraph 2 (shut the 
instrument off), but keep the 

ON/Zero

 button pushed 

until  the  wavelength  change  appears  on  the  display 
(this  step  takes  approximately  7  seconds).  At  this 

point release the button and the instrument is ready for a new measurement.. The 
displayed  value  is  the  last  recorded  power  measurement  carried  out  at  this 
wavelength setting.

 

11-Customer’s calibration

 

This  instrument  can  be  re-calibrated  by  the  user  to 
match a custom reference. Remove the plastic cover 
of  the adjustment switch located at the right side of 
the  instrument  body.  Zero  the  instrument  as 
indicated  in  Paragraph  2  of  this  Operation  Section 

and afterwards carry out a measurement of a known laser source. Shift the switch 
on + or – if you need to increase or decrease the reading displayed on the LCD to 
match your reference. This is done by limited pushes on the 

ON/Zero

 key, until 

the desired power value is achieved on the display. Reset the switch to its middle 
position 

CAL

 and re-cap. The instrument is ready for new measurements.  

 

 

 

Summary of Contents for Cronos Series

Page 1: ...er of this instrument must be trained to the use of power lasers and associated risks LaserPoint is in no way liable for any damage resulting from improper careless or above rated limits use of this i...

Page 2: ...art 1 Residential commercial and light industrial EN 50082 2 1995 Immunity generic standard Part 1 Industrial Standard Description Performance Criteria EN 55022 1994 amendment 1 1995 amendment 2 1997...

Page 3: ...time of calibration the above listed instrument meets or exceeds all technical specifications defined at page 8 of this Instruction Manual The above listed instrument has been calibrated using standar...

Page 4: ...ur instrument LP X where X 1 5 5 0 or 10 then release it this step takes approximately 2 seconds The instrument will also display the selected laser wavelength two options are possible CO2 10 6 m or Y...

Page 5: ...Display As soon as the measurement is over the LED turns into steady red Remove the probe from the laser beam the display shows the actual power and the remaining thermal capacity If the bar graph is...

Page 6: ...f but keep the ON Zero button pushed until the wavelength change appears on the display this step takes approximately 7 seconds At this point release the button and the instrument is ready for a new m...

Page 7: ...item without cost to yourself It is emphasised that this no cost return option is only available when you purchase a new product of equivalent type that fulfils the same function Please inform your su...

Page 8: ...ontinuous operation without battery replacement h 200 Operating temperature range C 10 to 40 Storage temperature range C 10 to 60 Weight g 480 950 1300 Length with handle mm 300 310 320 Specifications...

Reviews: