ICS-6-OEM Instruction Manual
01807-53-000
Page 25 of 32
Issue 5 19 January 2021
A suitable tool to reset the Mismatch button includes a small bladed screwdriver or similar.
8.3 Mismatch Detector Remote Test and Reset
Terminals are provided that allow for remote testing and resetting of the mismatch detector if
desired. For example, a key-operated switch could be connected to the terminals so that when it is
set one way it tests the mismatch detector, and when set the other way resets it. The key is then
returned to the mid position and removed to allow normal operation to resume.
The ‘test’ switch should have a contact which is normally closed, and which opens to perform the
test. The ‘reset’ switch should have a contact which closes to reset the detector.
If the test function is not used a wire link must be fitted across the ‘TST’ terminals.