Average / Standard deviation
For several measurements, the average
_
x indicates the average value, while
the standard deviation (Sdev) gauges the average deviation of the individual
LD@RTQDLDMSREQNLSGHR@UDQ@FDU@KTD,NQDRHFMHÖB@MSRS@MC@QCCDUH@SHNMR
indicate a more scattered series of measurements.
For normal measurement distributions,
68% of the measurements are within
_
x ± (1*Sdev),
95% of the measurements are within
_
x ± (2*Sdev),
99% of the measurements are within
_
x ± (3*Sdev)
Error messages
Error code
Description
Err1, Err2, Err3
Sensor not connected properly.
Deviating signal.
Err 1
Error in eddy current sensor
Err 2
Error in magnetic induction sensor
Err 3
Error in both sensors
Err 4, Err 5, Err 6
reserved
Err 7
Error in coating thickness
If errors consistently recur, please contact your specialist dealer
or the Laserliner service personnel.
!
Notes:
– The statistics data must be deleted before each new measuring task
ÖQRSLD@RTQDLDMSMN
– In the case of rough, contaminated surfaces, averaging may improve
the measurement result.
– Statistics can be used to determine the average coating thickness on
extensive measurement areas; to do this, set at least 10-20 measuring
points distributed across the entire area.