
26
10
10.1
11
Zero
μ
m
DIR
Fe
NO
AVG 100
101.0
1
The zero point setting should be checked and, if necessary, corrected before
each measurement.
Zero calibration
Checking/Calibrating the zero point
1
Turn the device off/on
2
Set the measuring mode:
Menu > Options > Measure mode > Single mode
Exit the menu: press key 1 twice
3
Perform measurement (see item 12):
measurement on uncoated material
(original material or reference material)
Value within tolerance
(see technical data):
check complete; no calibration necessary
Value outside tolerance
(see technical data):
Calibration required: press and hold the ZERO key (3);
repeat steps 1-3 to check
One-point calibration calibrates a coating thickness as close
to the expected measured value as possible.
One-point and two-point calibration
3VNONHMSB@KHAQ@SHNMCDÖMDR@RDBNMCBN@SHMFSGHBJMDRR
in order to increase the measuring accuracy. The expected
measured value should be as close as possible to the mid-
point between the calibration points (cal 1 and cal 2).
,TKSHOKDÖKLRL@XADBNLAHMDCENQFQD@SDQBN@SHMF
thicknesses.
Example:
expected coating thickness:
75
μ
m
Calibration point 1:
,D@RTQDLDMSVHSGQDEDQDMBDÖKL
50
μ
m ± 1%
Calibration point 2:
,D@RTQDLDMSVHSGQDEDQDMBDÖKL
101
μ
m ± 1%
EN