EPS2116 Reference Manual
Calibration
3-1
Module 3
Operations
3.1
Calibration ...........................................................................................................................3-1
3.2
Windscreen Replacement ...................................................................................................3-3
3.3
Desiccant Replacement ......................................................................................................3-5
3.1 Calibration
To calibrate microphones in the EPS2116, follow these steps:
TAKE NOTE
The microphone
and preamplifier can be
pulled up if needed to allow
the microphone to seat fully
inside the calibrator, though
this is not necessary when
using the CAL200 or CAL250.
Step 1
Remove birdspike and windscreen by unscrewing the
assembly.
FIGURE 3-1
Remove birdspike
Step 2
Place the calibrator (CAL250 shown) on the microphone by
pushing down gently while slightly twisting clockwise.
Summary of Contents for EPS2116
Page 1: ...EPS2116 Outdoor Microphone Preamplifier Protection...
Page 2: ...Larson Davis EPS2116 Environmental Protection System Reference Manual...
Page 19: ...EPS2116 Reference Manual A 3 A 5 EPS2116 Dimensions FIGURE A 1 EPS2116 Dimensions...
Page 27: ...EPS2116 Reference Manual A 11...
Page 28: ...EPS2116 Reference Manual A 12...
Page 29: ...EPS2116 Reference Manual A 13...
Page 30: ...EPS2116 Reference Manual A 14...
Page 32: ...EPS2116 Reference Manual A 16...
Page 33: ...EPS2116 Reference Manual A 17...
Page 34: ...EPS2116 Reference Manual A 18...
Page 36: ...EPS2116 Reference Manual A 20...
Page 37: ...EPS2116 Reference Manual A 21...
Page 38: ...EPS2116 Reference Manual A 22...
Page 39: ...EPS2116 Reference Manual A 23...
Page 40: ...EPS2116 Reference Manual A 24...