Page 14 / 23
No
Function
GET_LASER
2
Remark
1) Read the set value of laser
Byte
1
2
3
4
5
6~N
N+1
N+2
STX
ID
CMD
SIZE
DATA
CS
ETX
Command
Packet
05h
SYS ID
HOST ID
E1h
TS1
TD1-TDN
CS
06h
Ack Packet
05h
HOST ID
SYS ID
E1h
RS1
RD1-RDN
CS
06h
Parameter
TS1, TD1~TDN: Please refer to the following table
RS1, RD1~RDN: Please refer to the following table
Remark
TS1
TD1-TDN
01h
[TD1] LASER_MONITORING (01h): Request current set value
[TD1] LASER_FW_VERSION (02h): Request firmware version
[TD1] FAN_SENSOR_TEMP (52h): Get the sensing temperature of the fan sensor
RS1
RD1-RDN
06h
TD1=
LASER_MONITORING ,
[RD1] LASER_MONITORING (01h)
[RD2] Temp Value (00h ~ FFh)
[RD3] LD Out Value (00h ~ 64h)
[RD4] MOT Move Value (00h ~ FFh) Low Byte
[RD5] MOT Move Value (00h ~ 0Fh) High Byte
[RD6] Status Value (00h ~ FFh)
03h
TD1= LASER_FW_VERSION ,
[RD1] LASER_FW_VERSION (02h)
[RD2] App Version (00h~09h) High Byte
[RD3] App Version (00h~63h) Low Byte
TD1= FAN_SENSOR_TEMP ,
[RD1] FAN_SENSOR_TEMP (52h)
[RD2] Sensor Temp (00h~FFh) Low Byte (The value of multiple by 100 for two decimal places)
[RD3] Sensor Temp (00h~FFh) High Byte (The value of multiple by 100 for two decimal places)