8.3 ERRORS
During the processing, the following errors can occur. The user can repeat
processing. If the error persists the machine must be serviced by a author-
ized repair centre.
WATER FLOW TOO LOW ERROR
The water flow rate is low.
This can be due to:
- Air has accumulated in the system, perform
a CLEANING PROCEDURE, see maintenance
section below.
- Filters are clogged, replace the filters and per-
form a FILL FILTER procedure, see maintenance
section below
SCALE ERROR
Scales reading wrong values
BOILER ERROR
Boiler not reaching operating temperature.
BOILER ERROR
Boiler overheating.