SEM5000 Surface Emissions Monitor
ATEX WARNINGS
Read the entire SEM5000 User Manual prior to operating the instrument
This warning is issued to ensure proper knowledge of the instrument’s operation.
Replacment of the battery must only be performed in non-hazardous areas
Do not charge the battery in hazardous areas
The charging of the batteries must be performed only in a safe place with the appropriate
external adapter QED SEM-BC.
Do not open the instrument
Only authorized personnel can open the instrument for maintenance and repair.
The opening of the instrument by unauthorized personnel will void the product warranty.
Maximum voltage Um < 15V
The maximum voltage that can be applied to the non-intrinsically safe connection facilities of
associated apparatus (battery charger) without invalidating the type of protection is 15 volts.
Antistatic Personal Protection Equipment
The operator shall be equipped with a complete antistatic Personal Protection Equipment
(PPE), in combination with conductive or dissipative ground, and shoes with a resistance
below 10
8
Ω.
Do not connect the SEM5000 instrument, SEM5000 battery or SEM5000 battery charger
to other equipment
These devices are only designed to be used together in order to work properly and guarantee
the safety of the users.
Atex accessories
Use only the following accessories with the instrument SEM5000:
Ø
Rechargeable battery pack lithium ion (PBLO.NNNN.YY 3,7V 4000mAh)
Code: 205014
Ø
Battery charger (CCLO.NNNNN.YY)
Code: 100189
Ø
Power supply (Um<15 volts)
Code: 423007