SEM5000 Surface Emissions Monitor
Pump Error
The instrument detects when the pump flow is compliant, otherwise the pump stops, the
LEDs light up, the icon of the pump shows a cross and you will see the message "PUMP
ERROR" is displayed.
The instrument detects when the pump flow is restricted or blocked. The pump stops, the
LEDs light up, the icon of the pump shows a cross and you will see the message "PUMP
ERROR" is displayed on the screen.
If this error occurs, turn off the SEM5000, check the water trap filter and probe ends for
potential restrictions. Once you have cleared any restrictions, restart the instrument.
If you want to turn the pump off manually, press the Selector Dial, scroll to the "ON / OFF
Pump" option and press the dial again. Once pressed, the “Pump” icon will be displayed with
an “X”, indicating that the pump is off.
NOTE: This menu is available only when the GPS is activated and the “Save Location” icon is
shown on the display. When the pump is off, the display shows five horizontal lines “-----"
Restarting the Pump
•
Press the Function button
•
The “X” over the pump icon will disappear and the pump will start again
The field in the upper, left-hand corner of the screen can display two different icons:
When the GPS is off, the “Pump” icon is displayed.
When the GPS is on, the “Save Location” icon is displayed.
Note: For further information please contact Technical Support at QED at (800) 968-2026 or
email [email protected].