810-815 LX / INSTALLATION
NI_810-815_LX_PILOTE_ANG_A.doc
14
WATER CONNECTIONS
(INTERMEDIATE PRESSURE DEMINERALIZED SUPPLY / 01060206)
COLD WATER :
A valve with a male threaded nozzle, 20/27 diameter, (3/4" hose thread) is
required.
Flow rate : 20 l/min (5.3 GPM) with a minimum pressure of 200 kPa (2 bar or
29 PSI) and maximum 600 kPa (6 bar or 87 PSI).
DEMINERALIZED, DISTILLED OR REVERSE OSMOSIS WATER :
By stainless steel, PVC or similar piping.
A valve with a threaded male nozzle, 20/27 diameter (3/4" hose thread) is
required.
Flow rate: 20 l/min (5.3 GPM) under minimum pressure of 35 kPa (0.35 bar or 5
PSI) and maximum 600 kPa (6 bar or 87 PSI).
The waters intakes will have to be in waters intakes setting up area (
see drawing
next page
).
Remove the rear bodywork to accede to the connection support.
WATER CONNECTIONS
-1- COLD WATER
-2- DEMI. WATER