14
Auto Off
Yes, default 15 (only with battery power)
Power Save
Yes, default OFF
Data Logger
250 test results stored for download to PC or transfer by
Bluetooth
Certifications
EZ-BLE
™
PRoC
™
Module,
CYBLE-022001-00
RF Radio:
FCC (USA):
Industry
Canada (IC)
Certification:
FCC ID:
WAP2001
License IC:
7922A-2001
CE (Europe):
Complies with
Directive
1999/5/EC
MIC (Japan):
005-101007
KC (Korea):
MSIP-CRM-
Cyp-2001
EMC:
EU: ETSI EN 301489-1
US: FCC PART 15 B
CAN ICES-3 (B)/NMB-3(B)
AS/NZS: CSPR 22
Safety:
EU: EN61010-1:2010
AS/NZS: national differences
Dimensions
21.6 X 12.4 X 10.4 cm (L X W X H)
8.5 X 4.9 X 4.2 in
Weight
0.79 Kg, 1.74 lb
TROUBLESHOOTING
TROUBLESHOOTING GUIDE
Problem
Reason
Solution
“No Disk, Test Aborted”
No disk in chamber
Empty chamber
Place filled disk in chamber. Cover
with disk cover. Select OK. Start
test.
Insert filled disk. Cover with disk
cover. Select OK. Start test.
“No Disk Cover, Test
Aborted”
Disk cover was not used
Use disk cover. Select OK. Start
test.
“Lid Open, Test Aborted”
Lid opened
Close lid. Select OK. Start test.
“Used Disk”
Reacted disk in chamber
Select “Continue” to go to Test
Results screen. Select “Abort” to
go to Testing screen and run test
with new disk.