5. To cancel or abort this operation, press the
CAL
key. Note no new value
will be stored into the meter's non-volatile memory. To confirm the
calibration, press the
ENTER
key. The LCD displays “
CO
” momentarily,
and the meter reverts to the measurement mode.
3.4.2 Without Temperature probe (no ATC)
If a temperature probe is not used, the meter compensates for the pH response
based on a new calibrated temperature value manually set or at 25.0 °C
(factory default).
1.
Press the
MODE
key to enter into Temperature mode until “°
C
” shows on
the LCD.
2.
Compare the displayed value to a NIST certified thermometer or
thermometer known to be accurate (dipped into a constant temperature
bath).
3.
Press the
CAL
key to enter the temperature calibration mode. The LCD
shows “
CA
” momentarily and the displayed reading flashes. Note that this
displayed value should either be 25.0
o
C or the last set temperature value.
4.
Press
p
and
q
keys until the display shows the desired temperature. Any
value from 0 to 100
o
C can be set.
5.
To cancel or abort this operation, press the
CAL
key. Note no new value
will be stored into the meter‘s non-volatile memory. To confirm the
calibration, press the
ENTER
key. The LCD displays “
CO
” momentarily,
and the meter reverts to measurement mode.
18