TROUBLESHOOTING

Problem

Cause

Action

Power on but no display Batteries

Insert batteries

Batteries

Verify correct polarity

Batteries

Replace batteries

Unstable readings

Electrode

Immerse electrode more deeply 

in sample

Electrode

Condition electrode before f rst 

use. Soak in tap water for 10 

minutes.

Electrode

Remove air bubbles caught 

under electrode 

Electrode

Clean electrode

Electrode

Replace electrode

Static charge

Wiping electrode Rinse electrode and blot. Do not 

wipe electrode

Same readings in 

different  samples

Cracked or broken 

electrode

Replace electrode. Avoid 

plunging electrode into bottom 

of container and stir bars.

Display frozen

HOLD function

Press MODE or turn meter off

Button response

Remove batteries (stored data 

will be lost)

Steady “-1 ” display

Wait

Reading not stable yet

REPLACING THE ELECTRODE

The TRACER is shipped with an electrode attached. If the electrode needs to be 

replaced, follow these steps for removing and connecting electrodes.
1. Turn meter off.
2. To remove the electrode, turn the collar counter-clockwise and remove it.
3. Gently rock the electrode from side to side, pulling it downwards, until it

disconnects from the meter.

4. To attach an electrode, carefully plug the electrode into the meter socket.

CAUTION: Take care to align the pins carefully. Bent or broken pins will

cause the meter to malfunction.

5. Tighten the electrode collar f rmly enough to make a good seal. A rubber

gasket will seal the electrode with the meter.

Summary of Contents for ORP PockeTester Tracer

Page 1: ...Water Testing Leader Since 1919 Tracer ORP PockeTesterTM Code 1742 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Page 2: ...Display 5 Basic Operation Powering the TRACER 6 Electrode Recognition 6 Automatic Calibration 6 Data Hold 6 ORP Testing Overview 6 ORP Display 6 Getting Started 6 ORP Measurement 6 Storing Readings 7...

Page 3: ...unction display with bar graph Operating conditions 32 to 122 F 0 to 50 C and 80 RH ORP Range Accuracy 999mV 4mV Measurement storage 15 readings numbered readings Low battery indicator BAT appears on...

Page 4: ...ver 2 LCD Display 3 MODE HOLD button 4 CAL RECALL button 5 ON OFF button 6 Electrode Collar 7 Electrode Sensor Note The Electrode storage cap is not shown LCD Display 1 Reading 2 Units 3 Bargraph 4 Da...

Page 5: ...Hold Press the MODE HOLD button to freeze the current reading The HOLD display icon will appear along with the held reading The held reading will be stored in the next 15 reading storage location Pre...

Page 6: ...1 to store the next reading 7 If an attempt is made to store more than 15 readings the stored readings starting with the f rst reading will be overwritten Recalling Stored Readings NOTE First ensure t...

Page 7: ...Store vertically Battery Replacement 1 Twist off the battery compartment cap 2 Hold the battery housing in place with one f nger Remove the battery carrier by pulling on the small tabs 3 Replace the...

Page 8: ...ars Display frozen HOLD function Press MODE or turn meter off Button response Remove batteries stored data will be lost Steady 1 display Wait Reading not stable yet REPLACING THE ELECTRODE The TRACER...

Page 9: ...des Total Chlorine TRACER Electrode 0 00 10 00 ppm The Total Chlorine TRACER Electrode Code 1732 requires the use of TRACER TCL Tablets Order Code 7044A J 100 pack Code 1732 pH TRACER Electrode 0 00 1...

Page 10: ...of the user such as misuse improper wiring operation outside of specif cation improper maintenance or repair or unauthorized modif cation LaMotte Company specif cally disclaims any implied warranties...

Reviews: