LaMotte ColorQ DW Manual Download Page 8

800-344-3100 • 410-778-3100 • 410-778-6394

www.lamotte.com

PO Box 329 • Chestertown • Maryland • 21620 • USA

Color

Q

User Tips

!

Always rinse tubes and caps after testing.
Reagent left behind can affect the next test or
stain the tube.

Error Messages

Use a small Phillips head screwdriver to
remove all 5 screws in the base. Gently pull
the battery holder from the inside and
replace TWO AA style batteries. Replace
screws securely but do not over-tighten
screws which can tear the rubber base.

bAt

= Battery is low.

Maintenance

Battery Replacement

Hi

= Result is above the effective measuring

range for the system (e.g. >9.0 pH for Code
7059 reagent.)

!

Replace stained or scratched tubes (0201).

!

Clean ColorQ optics with a damp cotton
swab. Avoid abrasive cleaners that can
damage plastic.

Do not use

brush

the test tube

to clean

the ColorQ meter chamber.

!

Be sure the outside of the tubes are dry
before putting them into the ColorQ. This
is best accomplished by dispensing the
water sample from the sample bottle.

!

Replace liquid reagents annually. To learn
more about reagent storage guidelines and
shelf-life go to Support Section then Reagent
Refills/Shelf Life at www.lamotte.com.

Low pH

Low pH levels can significantly

lower the hardness results.

Er 3

= Insufficient light is reaching the

detector. Sample may be too dark, or tube
may be in sideways. Turn off meter and
retest.

Er 7

= The blank was not properly set. Turn

off meter and retest blank making sure the
black arrow on the tube is facing forward,
toward the user (not left or right).

SUN

= Extremely bright sunlight is

penetrating the chamber. Turn your back to
the sun to shade the meter from bright
sunlight.

Lo

= Result is below the effective measuring

range for the system (e.g. < 4.0 pH for
Code 7059 reagent.)

!

If not testing all tests in the sequence,
follow blanking procedure and then press
button until desired test factor is shown.
Follow test procedure.

!

In one complete inversion, the tube will be
turned cap down and then right side up.
The air bubble will move slowly to the
bottom of the tube and back again to the
cap end.

!

If small bubbles form after adding
reagents, tap bottom of tube sharply once
or twice to dislodge bubbles. Bubbles will
interfere with the test results.

!

Samples with concentrations greater than the
range of the reagent system must be diluted.
Add 2.5 mL of sample water to a test tube.
Fill to the 5 mL line with distilled or deionized
water. Cap and mix. Blank with the diluted
sample. Follow the test procedure with the
diluted sample. Multiply the displayed result
by 2. (This will not work for the pH test).

Testing Tips

!

Read all instructions before testing.

62059-01   11/19

Summary of Contents for ColorQ DW

Page 1: ...own Maryland 21620 USA bAt Battery is low Use only LaMotte replacement reagents If the unit should auto off during testing rinse and fill one tube to Blank and return to testing Insert tubes into Colo...

Page 2: ...s button to Blank the meter and go to pH Remove tube 1 Fill sample bottle with water sample Replace cap 2 Fill clean tube to the 5mL line with water sample 3 Insert tube into ColorQ as shown pH 2 pH C...

Page 3: ...ardness in gpg Remove tube 4 Cap Invert 3 times to mix Insert tube into ColorQ DILUTION A Use the 1 mL pipet to add 2 mL water sample to a clean test tube C Blank with the diluted sample from step B E...

Page 4: ...etely Insert tube into Color Q DILUTION Samples with iron concentrations above 3 ppm must be diluted The meter will display Hi A Use the 1 mL pipet to add 1 mL water sample to a test tube B Dilute to...

Page 5: ...ube 1 Press button to go to FCL Free Chlorine 3 Add 1 DPD 1 Tablet Cap Shake for TEN seconds Invert slowly 5 times 2 Fill a clean tube to the 5 mL line with water sample Free Chlorine and Total Chlori...

Page 6: ...Add 5 drops of SULFIDE REAGENT C 8 Press button to read SuL Sulfide in ppm Remove tube 1 Press button to go to SuL Sulfide DILUTION Samples with sulfide concentrations above 3 ppm must be diluted The...

Page 7: ...trogen in ppm Remove tube 4 Cap Invert 30 times per minute for TWO minutes 3 Add 1 Tablet Crush with tablet crusher Immediately slide tube into Protective Sleeve NITRATE NOTE Nitrate Tablets are sensi...

Page 8: ...ness results Er 3 Insufficient light is reaching the detector Sample may be too dark or tube may be in sideways Turn off meter and retest Er 7 The blank was not properly set Turn off meter and retest...

Reviews: