LaMotte ColorQ DW Manual Download Page 5

5

ColorQ Range: 0.0-10.0 ppm Chlorine

FREE CHLORINE / TOTAL CHLORINE

DPD 3 - Code 6197A-H

DPD 1 - Code 6903A-H

5

/

Fcl

/

+cl

4. Insert tube into ColorQ.

2. Remove cap from reacted FCL tube.

5. Press button to read “tCL” Total Chlorine in ppm.  Remove tube.

3. Add 1

Tablet. Cap. Shake for TEN seconds. Invert slowly 5 times.

DPD 3

1. Press button to go to “tCL” Total Chlorine.

4. Press button to read “FCL” Free Chlorine in ppm.  Remove tube.

1. Press button to go to “FCL” Free Chlorine.

3. Add 1 DPD 1 Tablet. Cap. Shake for TEN seconds. Invert slowly 5 times.

2. Fill a clean tube to the 5 mL line with water sample.

Free Chlorine and Total Chlorine

7. Press button to read “tCL” Total Chlorine in ppm. Remove tube.

6. Insert tube into ColorQ.

1. Press button past “FCL” Free Chlorine to go to “tCL” Total Chlorine.

3. Add 1

Tablet. Cap. Shake for TEN seconds. Invert slowly 5 times.

DPD 1

4. Remove cap. Add 1

Tablet. Cap. Shake for TEN seconds.

Invert slowly 5 times.

DPD 3

2. Fill a clean tube to the 5 mL line with water sample.

Total Chlorine Only

+ C L

Color

Q

1

+ C L

Color

Q

Summary of Contents for ColorQ DW

Page 1: ...own Maryland 21620 USA bAt Battery is low Use only LaMotte replacement reagents If the unit should auto off during testing rinse and fill one tube to Blank and return to testing Insert tubes into Colo...

Page 2: ...s button to Blank the meter and go to pH Remove tube 1 Fill sample bottle with water sample Replace cap 2 Fill clean tube to the 5mL line with water sample 3 Insert tube into ColorQ as shown pH 2 pH C...

Page 3: ...ardness in gpg Remove tube 4 Cap Invert 3 times to mix Insert tube into ColorQ DILUTION A Use the 1 mL pipet to add 2 mL water sample to a clean test tube C Blank with the diluted sample from step B E...

Page 4: ...etely Insert tube into Color Q DILUTION Samples with iron concentrations above 3 ppm must be diluted The meter will display Hi A Use the 1 mL pipet to add 1 mL water sample to a test tube B Dilute to...

Page 5: ...ube 1 Press button to go to FCL Free Chlorine 3 Add 1 DPD 1 Tablet Cap Shake for TEN seconds Invert slowly 5 times 2 Fill a clean tube to the 5 mL line with water sample Free Chlorine and Total Chlori...

Page 6: ...Add 5 drops of SULFIDE REAGENT C 8 Press button to read SuL Sulfide in ppm Remove tube 1 Press button to go to SuL Sulfide DILUTION Samples with sulfide concentrations above 3 ppm must be diluted The...

Page 7: ...trogen in ppm Remove tube 4 Cap Invert 30 times per minute for TWO minutes 3 Add 1 Tablet Crush with tablet crusher Immediately slide tube into Protective Sleeve NITRATE NOTE Nitrate Tablets are sensi...

Page 8: ...ness results Er 3 Insufficient light is reaching the detector Sample may be too dark or tube may be in sideways Turn off meter and retest Er 7 The blank was not properly set Turn off meter and retest...

Reviews: