*Nitrate - Code 3884A-H
7
1. Press button to go to “nit” Nitrate.
5. Wait FIVE minutes. Remove sleeve if used. Insert tube into Color Q.
2. Fill a clean tube to the 5 mL line with water sample.
6. Press button to read “nit” Nitrate-Nitrogen in ppm. Remove tube.
4. Cap. Invert 30 times per minute for TWO minutes.
3. Add 1
Tablet. Crush with tablet crusher. Immediately slide
tube into Protective Sleeve.
NITRATE
NOTE: Nitrate Tablets are sensitive to UV light. The Protective Sleeve
will protect the reaction from UV light. If testing indoors, there is no
need to use the Protective Sleeve in this procedure.
7
ColorQ Range: 0-15 ppm
NITRATE-NITROGEN
1
Color
Q
Color
Q