6. Press
ENTER
to select
Turbidity Calibration.
Turbidity Calibration
Formazin
Japan Standard
12:00:00 001/500
7. Scroll to the desired
calibration option. Select
a calibration option based
on the composition of the
standards that will be used to
calibrate the meter.
Turbidity Calibration
Formazin
Japan Standard
12:00:00 001/500
NOTE: Stablcal® standards below 50 NTU should not be used to
calibrate the 2020we/wi. The diluent has a different refractive index than
traditional formazin standards and will affect the results.
8. Press
ENTER
to save the
selection. The screen will
display
Storing... for about
1 second and return to the
Turbidity Options menu.
Press
EXIT
to return to a
previous menu.
Turbidity Options
Turbidity Units
Turbidity Calibration
12:00:00 001/500
23
Options/Set Up
Summary of Contents for 2020we
Page 1: ...2020we wi 1970 EPA 1970 ISO...
Page 2: ......
Page 57: ...57...
Page 58: ...58...
Page 59: ...59...