SAMPLE DILUTION TECHNIQUES
If a test result is out of the range of the meter, it must be diluted. The
test should then be repeated on the diluted sample. The following table
gives quick reference guidelines for dilutions of various proportions.
Amount of Sample Deionized Water to Bring
Final Volume to 10 mL
Multiplication Factor
10 mL
0 mL
1
5 mL
5 mL
2
2.5 mL
7.5 mL
4
1 mL
9 mL
10
0.5 mL
9.5 mL
20
All dilutions are based on a fi nal volume of 10 mL, so several dilutions
will require small volumes of the water sample. Graduated pipets should
be used for all dilutions. If volumetric glassware is not available, dilutions
can be made with the colorimeter tube. Fill the tube to the 10 mL line
with the sample and then transfer it to another container. Add 10 mL
volumes of deionized water to the container and mix. Transfer 10 mL of
the diluted sample to the colorimeter tube and follow the test procedure.
Repeat the dilution and testing procedures until the result falls within
the range of the calibration. Multiply the test result by the dilution factor.
For example, if 10 mL of the sample water is diluted with three 10 mL
volumes of deionized water, the dilution factor is four. The test result of
the diluted sample should be multiplied by four.
OPTIONS & SET UP
FACTORY DEFAULT SETTINGS
Settings that have user options have been set at the factory to default
settings.
The factory default settings are:
Averaging
Disabled
Turbidity Units
NTU
Turbidity Calibration
Formazin
Date Format
MM-DD-YYYY
Power Save
5 minutes
Backlight
10 seconds
Language
English
17
Options/Set Up