background image

OPERATING INSTRUCTIONS 
METEODIGIT IV (9164) 

 

 

 

 2  

1. OPERATING CONTROLS 

 
 
 

 

 
 
 
 
 
 
 
 
 
 
 

 

 
 
 

 

 
 
 

1

Measuring inputs (M0)  

 

 

for all LAMBRECHT Sensors 

 

2

Connection DC 12V 

 

 12V

DC

 

Mains adapter  

 

3

LCD-Anzeige 

 
 

a

 Function 

 

b

 

Measuring point, 2nd meas. value  

 

c

 

Units for 2nd measured value  

 

d

 

Units for 1st measured value  

 

e

 

1st measured value  

 

f

 

Operational states:  

 
 

 

Battery voltage < 3.6V 

 

 

Unlocked for adjust. purposes  

 

 

Measured value corrected  

 

 

Relative measuring  

 

 Operating keys 

 

 ON OFF 

 

Switch ON, to switch OFF: press long 

 

 

M

 

 

M

 

 

Meas. point selection  

 

 MAX 

,

 MIN 

    Max. / min. value, clear: press long 

 

 

MEM 

 

  Measured data storage, display value: press long 

 

 

CLR 

 

  Relative measuring, cancel: press long 

 

 

To switch ON with press keys: 
 

 

 

CLR 

 Reinitialization 

 

 

 MEM 

 

Device configuration  

 

 

M

 

 

Software version  

 

 MAX 

 u.

 MIN 

  

unlocking for adjustment  

 

Rear of device 

 

5

Battery compartment  

 

3 AA alkaline-manganese batteries 

 
 

 
 
 

Summary of Contents for METEODIGIT IV

Page 1: ...and cost effective solution e g for production facilities quality management air conditioning technology and ventilating applications on ships and in scientific laboratories Digital measuring techniq...

Page 2: ...tes Battery voltage 3 6V Unlocked for adjust purposes Measured value corrected Relative measuring Operating keys ON OFF Switch ON to switch OFF press long M M Meas point selection MAX MIN Max min valu...

Page 3: ...onitoring 6 7 2 Switching ON OFF reinitialization 6 8 CONNECTING THE SENSORS 6 8 1 1 Sensors 7 8 1 2 Measuring inputs and additional channels 7 9 DISPLAY AND KEYPAD 7 9 1 1 Display 7 9 1 2 Keypad 7 10...

Page 4: ...z The device is damaged by electrostatic dis charge or lightning z Improper handling respectively non observance of the operating instructions The manufacturer reserves the right to change the product...

Page 5: ...ot use the device inside detonation areas or filling stations 4 1 1 Special notes on use If the device is brought into the work room from a cold environment there is a risk that conden sation might fo...

Page 6: ...ement It is possible by setting the measured value to zero to perform relative measuring operations with respect to a reference value 6 INITIAL COMMISSIONING 1 Connect sensor to socket M0 1 2 Ensure t...

Page 7: ...igits b for depicting various measuring func tions a there are also 4 arrows f for depicting the operating status Function field Main field Display of measuring functions in the function field Measuri...

Page 8: ...d 55 Difference millivolt 1 mV d 26 Difference millivolt 2 mV d260 Difference volt V d2 60 Sensor voltage V UbAt Frequency Hz FrEq Function channels Maximum value of channel Mb1 Hi Minimum value of ch...

Page 9: ...CLR Select the next digit or entry is completed ON To terminate or cancel configuration MEM 11 1 Moving average METEODIGIT IV offers all the possibilities to calculate the average value In the first...

Page 10: ...e listed checks and reme dial steps the device still fails to behave as described in the operating instructions it must be returned to our factory in Goettingen with error description 13 DECLARATION O...

Page 11: ...ly not separated Housing L 127 x B 83 x H 42 mm ABS weight approx 260 g Suitable conditions Operating temperature 10 50 C storage temperature 20 60 C Ambient relative humidity 10 90 r h non condensati...

Reviews: