9
16131.5 ADDITIONAL SPECIFICATIONS
Measurand
Sensor body temperature
Measurand in SI radiometry units
temperature in °C
Temperature sensor
solid state System on Chip (SoC) silicon bandgap temperature
sensor
Instrument body temperature accuracy
0.5 °C
Instrument body temperature resolution
3.9 x 10
-3
°C
Levelling (see options)
bubble level and adjustable levelling feet are included
Levelling accuracy
< 0.1 ° bubble entirely in ring
Connector type
M12-A, 5-pole, IP67
Mounting
2 x M5 bolt at 46 mm distance on north-south axis
IP protection class
IP67
Net weight including 5m cable
0.65 kg
HEATING (00.16131.501030 and 00.16131.501000 only)
Heater operation
the heater is not necessarily switched on; recommended
operation is to activate the heater when the sun is below the
horizon
Heater voltage
12 VDC
Required heater power
1.5 W at 12 VDC (the heater is not necessarily active)
Heater resistance
95 Ω
Steady state zero offset caused by heating
0 to -8 W/m2
CALIBRATION
Calibration traceability
to WRR
Calibration hierarchy
from WRR through ISO 9846 and ISO 9847, applying a
correction to reference conditions
Calibration method
indoor calibration according to ISO 9847, Type IIc
Calibration uncertainty
< 1.8 % (k = 2)
Recommended recalibration interval
2 years
Reference conditions
20 °C, normal incidence solar radiation, horizontal mounting,
irradiance level 1000 W/m
2
Validity of calibration
based on experience the instrument sensitivity will not change
during storage. During use under exposure to solar radiation
the instrument “non-stability” specification is applicable.
MEASUREMENT ACCURACY AND RESOLUTION
Uncertainty of the measurement
statements about the overall measurement uncertainty can
only be made on an individual basis.
WMO estimate on achievable accuracy for
daily sums
5 % (WMO-No. 8, seventh edition 2008)
WMO estimate on achievable accuracy for
hourly sums
8 % (WMO-No. 8, seventh edition 2008)
Irradiance resolution
0.01 W/m
2
00.16131.501030 –
DIGITAL MODBUS RTU (RS-485)
Digital output
irradiance in W/m2
instrument body temperature in °C
Rated operating voltage range
5 to 30 VDC (recommended 12 VDC)
Power consumption
< 48 x 10
-3
W at 12 VDC (unheated)
Heater
included
Communication protocol
Modbus over 2-wire RS-485
half duplex
RS-485 transceiver type
2-wire, non-isolated
Transmission mode
RTU