Lake Shore Model 625 Superconducting MPS User’s Manual
5-44
Computer Interface Operation
TRIG
Trigger Output Current Setting Command
Input:
TRIG <I value> [term]
Format:
±
nn.nnnn
<I value>
Sets the value in which the output current will ramp to when the instrument is
triggered: 0.0000 –
±
60.1000A.
Remarks:
This command only sets up the trigger. Use the
*TRG
command or the GET (group execute trigger)
function to start a trigger over the interface.
TRIG?
Trigger Output Current Setting Query
Input:
TRIG? [term]
Returned:
<I value> [term]
Format:
±
nn.nnnn
(Refer to command for description)
XPGM
External Program Mode Command
Input:
XPGM <mode>[term]
Format:
n
<mode>
0 = Internal, 1 = External, 2 = Sum.
Example:
XPGM 1[term] –
Places the Model 625 into external program mode where the output current is set by
an external voltage.
XPGM?
External Program Mode Query
Input:
XPGM?[term]
Returned:
<mode>[term]
Format:
n
(Refer to command for description)