6.4.1 Interface Commands
119
|
www.lakeshore.com
KRDG?
Kelvin Reading Query
Input
KRDG? <input>[term]
Format
a
<input>
Specifies which input to query: A –B.
Returned
<kelvin value>[term]
Format
±nnnnnn
Remarks
Also see the RDGST? query.
LEDS
Front Panel LEDS Command
Input
LEDS <off/on>[term]
Format
n
<off/on>
0 = LEDs Off, 1 = LEDs On
Remarks
If set to 0, front panel LEDs will not be functional. Function can be used when display
brightness is a problem.
Example
LED 0[term]
turns all front panel LED functionality off.
LEDS?
Front Panel LEDS Query
Input
LEDS?[term]
Returned
<off/on> [term]
Format
n (refer to command for description)
LOCK
Front Panel Keyboard Lock Command
Input
LOCK <state>,<code>[term]
Format
n,nnn
<state>
0 = Unlocked, 1 = Locked
<code>
Specifies lock-out code. Valid entries are 000 –999.
Remarks
Locks out all front panel entries except pressing
ALL OFF
to immediately turn off all
heater outputs. Refer to section 4.7.
Example
LOCK 1,123[term]
enables keypad lock and sets the code to 123.
LOCK?
Front Panel Keyboard Lock Query
Input
LOCK?[term]
Returned
<state>,<code>[term]
Format
n,nnn (refer to command for description)
MDAT?
Minimum/Maximum Data Query
Input
MDAT? <input>[term]
Format
a
<input>
Specifies which input to query: A or B.
Returned
<min value>,<max value>[term]
Format
±nnnnnn,±nnnnnn
Remarks
Returns the minimum and maximum input data. Also see the RDGST? query.
MNMXRST
Minimum and Maximum Function Reset Command
Input
MNMXRST[term]
Remarks
Resets the minimum and maximum data for all inputs.