1.3 Model 335 Specifications
7
|
www.lakeshore.com
1.3 Model 335
Specifications
1.3.1 Input
Specifications
Input specifications
Sensor
temperature
coefficient
Input range
Excitation
current
Display
resolution
Measurement
resolution
Electronic
accuracy
(at 25 °C)
Measurement temperature
coefficient
Electronic
stability
1
Diode
Negative
0 V to 2.5 V
10 µA ±0.05%
2,3
100 µV
10 µV
±80 µV ±0.005%
of rdg
(10 µV + 0.0005% of rdg)/°C
±20 µV
0 V to 10 V
10 µA ±0.05%
2,3
1 mV
20 µV
±320 µV ±0.01%
of rdg
(20 µV + 0.0005% of rdg)/°C
±40 µV
PTC RTD
Positive
0
)
to 10
)
1 mA
4
1 m
)
0.2 m
)
±0.002
)
±0.01% of rdg
(0.01 m
)
+ 0.001% of rdg)/°C
±0.4 m
)
0
)
to 30
)
1 mA
4
1 m
)
0.2 m
)
±0.002
)
±0.01% of rdg
(0.03 m
)
+ 0.001% of rdg)/°C
±0.4 m
)
0
)
to 100
)
1 mA
4
10 m
)
2 m
)
±0.004
)
±0.01% of rdg
(0.1 m
)
+ 0.001% of rdg)/°C
±4 m
)
0
)
to 300
)
1 mA
4
10 m
)
2 m
)
±0.004
)
±0.01% of rdg
(0.3 m
)
+ 0.001% of rdg)/°C
±4 m
)
0
)
to 1 k
)
1 mA
4
100 m
)
20 m
)
±0.04
)
±0.02% of rdg
(1 m
)
+ 0.001% of rdg)/°C
±40 m
)
0
)
to 3 k
)
1 mA
4
100 m
)
20 m
)
±0.04
)
±0.02% of rdg
(3 m
)
+ 0.001% of rdg)/°C
±40 m
)
0
)
to 10 k
)
1 mA
4
1
)
200 m
)
±0.4
)
±0.02% of rdg
(10 m
)
+ 0.001% of rdg)/°C
±400 m
)
NTC RTD
10 mV
Negative
0
)
to 10
)
1 mA
4
1 m
)
0.15 m
)
±0.002
)
±0.06%
of rdg
(0.01 m
)
+ 0.001% of rdg)/°C
±0.3 m
)
0
)
to 30
)
300 µA
4
1 m
)
0.45 m
)
±0.002
)
±0.06%
of rdg
(0.03 m
)
+ 0.0015% of rdg)/°C
±0.9 m
)
0
)
to 100
)
100 µA
4
10 m
)
1.5 m
)
±0.01
)
±0.04%
of rdg
(0.1 m
)
+ 0.001% of rdg)/°C
±3 m
)
0
)
to 300
)
30 µA
4
10 m
)
4.5 m
)
±0.01
)
±0.04%
of rdg
(0.3 m
)
+ 0.0015% of rdg)/°C
±9 m
)
0
)
to 1 k
)
10 µA
4
100 m
)
15 m
)
+0.002%
of rdg
±0.1
)
±0.04%
of rdg
(1 m
)
+ 0.001% of rdg)/°C
±30 m
)
±0.004%
of rdg
0
)
to 3 k
)
3 µA
4
100 m
)
45 m
)
+0.002%
of rdg
±0.1
)
±0.04%
of rdg
(3 m
)
+ 0.0015% of rdg)/°C ±90 m
)
±0.004%
of rdg
0
)
to 10 k
)
1 µA
4
1
)
150 m
)
+0.002%
of rdg
±1.0
)
±0.04%
of rdg
(10 m
)
+ 0.001% of rdg)/°C
±300 m
)
±0.004% of rdg
0
)
to 30 k
)
300 nA
4
1
)
450 m
)
+0.002%
of rdg
±2.0
)
±0.04%
of rdg
(30 m
)
+ 0.0015% of rdg)/°C
±900 m
)
±0.004% of rdg
0
)
to 100 k
)
100 nA
4
10
)
1.5
)
+0.005% of
rdg
±10.0
)
±0.04%
of rdg
(100 m
)
+ 0.002% of rdg)/°C
±3
)
±0.01%
of rdg
Thermocouple
Positive
±50 mV
NA
1 µV
0.4µV
±1 µV ±0.05%
of rdg
5
(0.1 µV + 0.001% of rdg)/°C
±0.8µV
1
Control stability of the electronics only, in ideal thermal system
2
Current source error has negligible effect on measurement accuracy
3
Diode input excitation can be set to 1 mA
4
Current source error is removed during calibration
5
Accuracy specification does not include errors from room temperature compensation
TABLE 1-3