Lake Shore Model 218 Temperature Monitor User’s Manual
6-30
Remote Operation
SCAL
Generate SoftCal™ Curve
Input:
SCAL <std>, <dest>, <SN>, <T1 value>, <U1 value>, <T2 value>, <U2 value>, <T3
value>, <U3 value>
Returned:
Nothing.
Remarks:
<std>
Specifies the standard curve to generate a SoftCal™ from (
1
,
6,7
).
<dest>
Specifies the user curve to store the SoftCal™ curve (
21
–
28
).
<SN>
Specifies the curve serial number. Limited to 10 characters.
<T1 value>
Specifies first temperature point.
<U1 value>
Specifies first sensor units point.
<T2 value>
Specifies second temperature point.
<U2 value>
Specifies second sensor units point.
<T3 value>
Specifies third temperature point.
<U3 value>
Specifies third sensor units point.
Example: SCAL 1, 21, SC12345, 4.2, 1.6260, 77.32, 1.0205, 300.0, 0.5189[term] –
Generates a 3-
point SoftCal™ curve from DT-470 and saves it in user curve 21 with serial number
SC12345.
SRDG?
Query Sensor Units Reading for a Single Input or all Inputs
Input: SRDG?
<input>
Returned: <sensor units value>
. Format:
+nn.nnn[term]
Or if all units are queried:
<Input 1 Sensor Units Value>,<Input 2 Sensor Units Value>,<Input 3 Sensor Units Value>,
<Input 4 Sensor Units Value>,<Input 5 Sensor Units Value>,<Input 6 Sensor Units Value>,
<Input 7 Sensor Units Value>,<Input 8 Sensor Units Value>
.
Format:
+nn.nnn,+nn.nnn,+nn.nnn,+nn.nnn,+nn.nnn,+nn.nnn,+nn.nnn,+nn.nnn
Remarks:
Returns the Sensor Units reading for a single input or all inputs.
<input>
Specifies which
input(s) to query.
0
= all inputs,
1-8
= individual input.
NOTE:
Use
0
(all inputs) when reading
two or more inputs at the maximum update rate of 16 rdg/s.
.