CPR Targets
Category
Specification
Compression Depth Target
> 50 mm ( 2”)
Depth accuracy: ±10%
Compression Release Target Force
< 2.5 kg (5.5 lbs)
Force accuracy: +1.5 kg to - 2.0 kg
(+3.3 lbs to - 4.4 lbs)
Compression Rate Target
100 to 120/min ± 3/min
CPRmeter 2 [REF 801-002xx]
The CPRmeter 2 meets the performance requirements of
IEC 60601-1, 3rd edition.
Category
Specification
Dimensions
153 mm x 64 mm x 25 mm (6.0” x 2.5” x 1.0”)
Weight
163 g (5.7 oz) (excluding batteries)
Batteries
2 x 1.5V Size AAA (LR03)
Temperature
Transport and Storage: -20 °C to 70 °C (-4 °F to 158 °F)
Operation: 0 °C to 50 °C (32 °F to 122 °F)
Sealing
Meets IP55 rating as defined by standard ISO/IEC 60529
Relative Humidity
Transport: 5% to 95%
Storage: 5% to 75%
Operation: 5% to 95%
Atmospheric
Pressure (Atm.p.)
Transport, Storage and Operation:
1014 to 617 mbar (1014 to 617 hPa)
Electromagnetic
Compatiblity
Meets IEC 60601-1-2 and RTCA/DO-160F Section
21 Category M
Defibrillation
Protection
The CPRmeter 2 is defibrillation protected, type BF
patient connection.
Specifications
26
Summary of Contents for CPRmeter 2
Page 1: ...CPRmeter 2 User Guide www laerdal com EN ES FR DE...
Page 2: ......
Page 6: ...A A A Items Included X 2 AAA Batteries CPRmeter 2 6...
Page 8: ...Overview On Off Status Light Compression Area Display Area Membrane Port 8...
Page 9: ...Overview Battery Hatch Placement Guide for Adhesive Placement Guide for Adhesive 9...
Page 10: ...x 2 4 3 2 1 Setup Insert Batteries 10...
Page 42: ...A A A Elementos incluidos X 2 Bater as AAA CPRmeter 2 42...
Page 46: ...x 2 4 3 2 1 Instalaci n Insertar bater as 46...
Page 78: ...A A A l ments inclus X 2 Piles AAA CPRmeter 2 78...
Page 82: ...x 2 4 3 2 1 Pr paration Insertion des piles 82...
Page 114: ...A A A Lieferumfang X 2 AAA Batterien CPRmeter 2 114...
Page 116: ...bersicht Ein Aus Statuslicht Kompressions bereich Anzeige bereich Membran port 116...
Page 118: ...x 2 4 3 2 1 Einrichtung Einlegen der Batterien 118...
Page 147: ......