8 MAINTENANCE AND SERVICE
LACO Technologies
TITAN VERSA Operations and Maintenance Manual
161
Pump Repair Options
There are three pump repair options summarized in Table 62 below.
Horizontal Wet Pump Repair Options
Maintenance Task
Part Number
Maintenance Interval
Level Site
Change pump oil
LVOEZUNO6
1500 h or 6 months
2
OS
Change oil mist eliminator filter PFPKE07025T
3000 h or 18 months
2
OS
Repair or exchange pump
LS-TVT-1 (Repair)
LS-TVT-2C (Exchange)
LS-TVT-2L (Exchange)
24000 h or 48 months 2-3
OS or LS
Replace pump
PFPKD07711
36000 h or 72 months 2-3
OS or LS
Other Pump Parts
VERSA L Pump Parts
Part Number
Description
LMSA119448
1015 hose assembly
LMK-TVL-5
1015 oil drain line
LMK-TVL-6
1015 oil fill line
LMK-TVL-7
1015 pump mount kit
CAUTION:
Dry pumps do not typically have short term maintenance
concerns but can potentially require major long-term repair if two-
year maintenance is not performed. Customers with leak detectors
in critical applications should consider purchasing a spare pump for
unplanned maintenance.
Removing the ISP-90 Pump
1.
Remove the back cover using 6 mm Hex wrench.
2.
Disconnect the Fore line hose from the pump.
3.
Disconnect the black exhaust hose from the chassis exhaust port.
4.
Unplug the power cord.
5.
Place leak detector on blocks to gain access to the bottom of the unit.
6.
Remove the four pump mounting bolts with a 5 mm hex wrench.
7.
Take the pump off for repair or maintenance.