Model KVVB Instructions 

Page 2 of 5 

 
 
4)  Operation: 

 
It is assumed that the user is familiar with the safety and cautionary practices required with high 

voltage systems and devices. Always exercise extreme caution when working with high voltages. 
 

WARNING: ALWAYS INSURE THAT HIGH VOLTAGE SOURCE IS OFF AND GROUNDED BEFORE 
CONNECTING. INJURY OR DEATH CAN RESULT FROM HIGH VOLTAGE. 

 
To use the divider, connect the green Ground binding post to a secure earth ground common 

with the high voltage source. Connect the high voltage input lead wire terminated in a #10 ring terminal 
to the high voltage source. Connect the Low and Output binding posts of the KVVB to the meter or 
measuring system. Note that output Low is at Ground potential. See figure 1. 

 

The KVVB divider may be operated in any position. Four corner mounting holes can be provided 

to accept #10 or M4 screws. 
 

To determine the input voltage from the measured output, use the following formula: 

 
Volts Input = (V Output) x (Divider Ratio) 
 

Example (for KVVB-10-1): 

Output  

0.775 V 

Ratio   

10,000:1 

Input   

0.775 x 10,000 = 7,750 Volts 

 

Care should be taken not to exceed the rated input voltage for the KVVB Divider. Note that the 

low terminal is internally connected to ground. 
 
 
 

 
 
 
 
 
 
 
 

 
 
 

Figure 1 

HV dc 
power 

source 

Load 

 
 

Input 

 
 

Gd   Low  Out 

>10 G voltmeter 