LABORIE AQUARIUS XT/CT/LT/CTS Service Manual
LBL-000233
[B]
38
The Urocap™ V should be cleaned using a damp cloth with alcohol, soap, or disinfectant detergent (Figure
25). Store the Urocap™ V in a cool and dry area at room temperature.
Figure 25: Cleaning the Urocap™ V
Rinse and dry the Uroflow beaker after each use. Use Soap and water, or a mild detergent solution, for cleaning.
Wipe the chair, Flowstand, and funnel clean with a cloth dampened with soap or mild detergent and wiped dry
immediately. When cleaning the funnel, the maximum cleaning temperature should not exceed 80°C (176°F).
To check if the Urocap™ V is functioning, see page 44.
NOTE: Although the beaker is reusable, it will discolor over time and may need to be replaced.
5.4
CARING FOR THE SPINNING DISK FLOW TRANSDUCER
On a weekly basis, test the device for proper functionality. Inspect all cables and the sensor for cuts and other damage. If
in doubt, send the device for service. No additional maintenance is required.
Clean the Flow Transducer at minimum once a week, preferably daily, to avoid deposits and odor. Clean the device as
follows:
Quick Cleaning (Daily)
Slowly pour approximately 1L of clean water into the funnel.
Keep the disc running to flush the transducer cup.
Thorough Cleaning (Weekly)
1.
Disconnect the Flow Transducer from the System.
2.
Disconnect the jug, the funnel, and the disk from the Spinning Flow Transducer.
3.
Use cleaning detergents appropriate for lavatory pans. Clean all the parts according to the detergent
manufacturers’ recommendations, including the inside and outside of the transducer cup.
4.
Wipe off all parts with an appropriate disinfection solvent.
IMPORTANT: To avoid spill, do not tilt the Flow Transducer during cleaning with liquid in the transducer cup.
5.
To avoid damage to the device, do not exceed the recommended cleaning temperatures for the different device
components:
Funnel
Max 140°C / 284°F
Rotating Disk
Max 150°C / 302°F
Transducer cup
Max 100°C / 212°F
Disc Holder
Max 100°C / 212°F
Jug 1.5L, Steel
Max 200°C / 392°F
Jug 1L, White
Max 140°C / 284°F
Motor Housing
Max 100°C / 212°F
Table 17: Component Cleaning - Fluid Temperatures
6.
After cleaning, rinse the components.
7.
Reassemble the flow transducer.
IMPORTANT:
Never immerse the connector into any liquid. Keep it dry at all times.
Never remove the disk holder for cleaning.