LABORIE AQUARIUS XT/CT/LT/CTS Service Manual
LBL-000233
[B]
30
4.2
HOW TO CALIBRATE THE PRESSURE AND THE EMG
Calibrate the pressure and EMG channels through the Roam DX. Make sure the transducer is connected to the device and
plugged in securely.
4.2.1
C
ALIBRATE
P
RESSURE
Table 11: Calibrating Roam™ DX Pressure Channels and Transducers
1.
Tap the Calibration icon on the home screen:
.
2.
Tap to select the Roam™ DX
device.
3.
Tap to select the Catheter System set up for calibration:
.
4.
The calibration window for the selected Catheter System will
load; in this example Water-based Disposable Catheter
System has been selected. Tap the boxes correlating with the
the transducer Channels being calibrated and tap Next to
continue:
NOTE: Scroll down the screen to see the other checkboxes (P4, P5)
.
5.
Type 0 (zero) in the low load box and tap Next:
.
6.
Continue calibration by following the on-screen instructions
until the calibration complete window appears:
.
7.
Tap the Return button to go back to the
Calibrate Pressure
Transducers screen:
8.
Repeat steps 2 to 7 for the
other Transducer Types.
.