 11.  

 INSTRUCTIONS FOR USING THE MICROSCOPE

Prima C/ Prima CS COLPOSCOPE 

a.     ADJUSTMENT OF INTERPUPILLARY DISTANCE (Fig. 2.13)

     

Turn the microscope in working position.

     Inclined tube 45°(1) : Adjust eyepieces (2) to the required IPD by turning the lateral knob (3).

     Straight tube (4) : Adjust eyepieces (5) by moving the binocular tube as per your requirement.

1

2

3

4

5

(Fig. 2.13)

22

6128000-795 

 Prima C/ Prima CS 

Issue 2.2
Printed on November,2020

Summary of Contents for Prima C

Page 1: ...S Federal Law restricts this device to sale by or on the order of a licensed Physician Rx only To ensure proper use of this instrument as well as to avoid injury while operating Instrument understandi...

Page 2: ...ma C Operating Instructions Prima C 9 1 On Off Switch 9 2 Brightness control Control elements Prima C CS Instructions for using Microscope a Adjustment of Interpupillary Distance b Changing the magnif...

Page 3: ...ABOMED dealer or from the LABOMED service department Contact information is provided at the end of this guide LABOMED Prima C CS COLPOSCOPE is an optical instrument used in a medical diagnostic proced...

Page 4: ...ructed in the expert handling of the device The proper installation and maintenance of the device are vital prerequisites before using the colposcope The user should check the functional stability and...

Page 5: ...hese operating instructions These instructions are structured around the operation of the colposcope from assembly to the finished system Basic troubleshooting information is also provided Our goal is...

Page 6: ...ification Knob Tilt Locking Knob CMO Focus Knob Eyepieces Binocular Head Binocular Head Locking Knob Arm Locking Knob Joystick Intensity Control Knob Prima C Ergonomic stand for easy and accurate posi...

Page 7: ...CMO Fine Focusing Knob Magnification Knob Tilt Locking knob Suspension arm hydraulic movement knob Suspension arm spring Tension Adjustment Prima CS Colposcope can be positioned easily with the help...

Page 8: ...D APPROPRIATE INSPECTION AND TESTING MUST BE CONDUCTED TO ENSURE CONTINUED SAFE USE OF THIS INSTRUMENT WARNING ENSURE THAT THE VOLTAGE APPLIED TO THE UNIT IS THE SAME AS THE VOLTAGE THAT IS INDICATED...

Page 9: ...CE SECTION FOR DETAILED CLEANING INSTRUCTIONS CAUTION MEDICAL ELECTRONIC EQUIPMENT NEEDS SPECIAL PRECAUTIONS WITH RESPECT TO ELECTROMAGNETIC CHARGE EMC AND NEEDS TO BE INSTALLED AND SERVICED ACCORDING...

Page 10: ...way from rain Connection ON OFF Year of manufacture used on PRODUCT DATA PLATE Alternating Current Power Fragile content of the transport package are fragile and therefore shall be handled with care C...

Page 11: ...Optics and Optical instruments Medical endoscopes and endoscopic accessories Part 3 Determination of field of view and direction of view of endoscopes with optics ISO 8600 5 First edition 2005 03 15 O...

Page 12: ...power supply Refer Fig 2 1 Page 13 Microscope Arm Refer Fig 2 2 Page 13 Microscope Head with Eyepieces Inclined or Straight Refer Fig 2 13 Page 21 Fasteners are enclosed in the packing box PRIMA CS T...

Page 13: ...TING OF ROLLER STAND AND COLUMN Fig 1 Insert column 1 into the mobile base 2 Fasten screw 3 with enclosed hexagon socket wrench to column 1 The mobile bases have 5 rollers 4 out of two have locking br...

Page 14: ...ation box 3 to the column against the vertical guide 2 Screws 1 to be tighten to hold the box as shown Fig 2 1 Insert the arm 4 of the microscope head on the top of the column 5 Screws 6 to be tighten...

Page 15: ...provided from the bottom of metal plate Refer Fig 2 3 Install the column D on the base Engage the column to the indexing screw E Align the tapped hole of the Column in the screw seat hold the column...

Page 16: ...screw driver in the socket situated in Suspension arm front coupling as shown as 7 in fig 2 5 The same is hanged with a tie to Potentiometer knob 5 1 SPECIAL INSTRUCTION The joint which connects swive...

Page 17: ...olposcope with H Base Stand Total weight Approx 86 5kg Three Boxes containing Base left wheel support 20 kg Base right wheel support 20 kg Base center Support 20Kg One Box Containing Column 8Kg Micros...

Page 18: ...w Instruction on electrical label provided at bottom of the arm as shown 3 here in Fig i 8 1 REPLACING THE ILLUMINATION SOURCE Open the swivel arm cover s Detach the fibre optic cable and replace the...

Page 19: ...Prima C Prima CS COLPOSCOPE 17 6128000 795 Prima C Prima CS Issue 2 2 Printed onNovember 2020 8 3 WIRING CODING DIAGRAM...

Page 20: ...as shown in figure number 2 8 below a CHANGING OF FUSE Note Fuses are in fuse holder F 2 8 2 main input fuses for lamp supply for mains voltage 100 240 V Changing of fuse Draw out the plug Insert a sc...

Page 21: ...n box After the appliance is swtiched on the green illumination switch diode indicates that the microscope is ready for operation The LED starts burning and the cooling fan starts working 9 2 BRIGHTNE...

Page 22: ...knob to change the Magnification 1 Common Main Objective CMO 2 Fine focusing knob 3 Joystick to tilt the head arm forward and backward 4 Head arm locking knob 5 Knob to lock the arm 6 Knob to swing in...

Page 23: ...ic movement knob 11 Magnification Knob 19 Suspension arm spring Tension Adjustment 12 Fine Focusing Knob 20 Microscope arm locking knob 13 Head Locking Knob 21 Suspension arm locking knob 14 Swivel ar...

Page 24: ...Fig 2 13 Turn the microscope in working position Inclined tube 45 1 Adjust eyepieces 2 to the required IPD by turning the lateral knob 3 Straight tube 4 Adjust eyepieces 5 by moving the binocular tube...

Page 25: ...N Fig 2 14 Adjust to the highest magnification with any of the rotating knobs 1 on the magnification changer See to it that the magnification step changer is engaged into its index position related to...

Page 26: ...mplete microscope in the focus range by any of the following means a Roll the mobile stand forward and backward b Adjust the arm towards left right and up down c Adjust the microscope angle left and r...

Page 27: ...ON FIG 2 16 Swing filter knob 1 to swing in or swing out the red free Green filter This will help in filtering the red tissue in the image being observed and will highlight the enhance vessel image an...

Page 28: ...NS FOR USING THE MICROSCOPE continued e EYEPIECES WITH DIOPTER Fig 2 17 You can adjust the diopter settings as per your vision correction Diopter adjustment is 5D Fig 2 17 26 6128000 795 Prima C Prima...

Page 29: ...Fig 2 18 The objective 1 has screw type mounting Unlock it by turning anti clock wise and lock by turning clock wise Eyepieces 2 are insert able Pull out for changing Other objectives eyepieces can be...

Page 30: ...straight or 45 inclined 2 Total four accessories are provided as follows 2 1 DSLR camera adapter for Nikon Canon and Sony 2 2 CCD camera adapter for Mintron 2 3 Adapter for Sony Handy Cam 2 4 Assisto...

Page 31: ...unscrew optical system and Mechanical parts unless explicitly instructed to do so in this manual Protect the microscope from oil and grease Never oil or grease the guide surfaces or mechanical parts T...

Page 32: ...d store them in a clean environment only Store under UV light when not in use Use in continuously climate controlled rooms only Keep moisture away using silica gel and cover with a plastic cover OCCUP...

Page 33: ...st it to feel wet Disinfect clean the surface gently Note 1 Use mask and Gloves while performing cleaning and disinfection 2 While cleaning muslin cloth should not be dripped wet to prevent seepage an...

Page 34: ...A C Prima C Prima CS COLPOSCOPE Schedule of Autoclavable caps Prima C Autoclavable caps C1 Part no 6122015 209 C2 Part no 6168000 219 C3 C4 Part no 6168000 213 32 6128000 795 Prima C Prima CS Issue 2...

Page 35: ...Autoclavable caps Prima CS Autoclavable caps C5 C6 C8 Part no 6122015 209 C7 Part no 6168000 219 C9 C10 C11 C12 Part no 6168000 213 Prima C Prima CS COLPOSCOPE 33 6128000 795 Prima C Prima CS Issue 2...

Page 36: ...s Prima C Prima CS COLPOSCOPE 34 6128000 795 Prima C Prima CS Issue 2 2 Printed onNovember 2020 For Operation For transportation and storage Temperature Rel humidity without condensation Air Pressure...

Page 37: ...Use constant voltage stabilizer to avoid sudden input voltage variation and to enhance LED life PRIMA CS H Base Stand Column lockable wheels Straight Observation tube interpupillary distance 50 75mm O...

Page 38: ...hone 510 445 1257 Fax 510 991 9862 Email sales laboamerica com www laboamerica com EU REP A16372 LABOMED ISO 13485 Labomed Europe Essebaan 50 NL 2908 LK Capelle a d IJssel The Netherlands Tel 31 0 10...

Reviews: