8B.
ELECTRICAL CONNECTIONS PRIMA C
Prima C/ Prima CS COLPOSCOPE
Connect the power cable to the AC Inlet socket as 1 in figure number 2.8 below.
Switch on the power from on/off switch shown as 1 in figure number 2.9 on page no. 18.
Note:
Power supply is designed with universal input 100V-240V AC, 50/60Hz. To plug in follow
Instruction on electrical label provided at back of illumination box as shown in figure number 2.8 below.
a. CHANGING OF FUSE
•
Note:
Fuses are in fuse holder
:
F 2.8(2 main input fuses for lamp supply for mains voltage 100-240 V)
•
Changing of fuse: Draw out the plug. Insert a screw driver into the slit of fuse holder and screw out to
the left. Remove cap and replace the fuse placed in it. Once the fuse is replaced cap it again and screw
in to the right. Refer electronical label for correct fuse replacement.
•
Attention: Fuses are allowed to change against the specified type only.
Fuse Inside
18
(Fig. 2.8)
.
6128000-795
Prima C/ Prima CS
Issue 2.2
Printed on November,2020