![Labomed Lx 400 User Manual Download Page 16](http://html1.mh-extra.com/html/labomed/lx-400/lx-400_user-manual_3214734016.webp)
Each standard set can be assembled by simply attaching the bulb, filter and power cord.
13
Lx 400
Fig. 15
Fig. 16
3
2
1
1
2
6
3
4
5
7
9
8
5
INSTALLING OR REPLACING THE LAMP BULB
PROCEDURE FOR LED REPLACEMENT
ASSEMBLY
9
1
2
Applicable LED replacement: LED P/N 9135000-901
1. Screws SS 658
2. Base Plate 9126100-301
3. Screws Heat Shield SS 714
4. Heat Shield 9135000-524
5. Star LED 9135000-901
6. LED Screws SS 658
7. Stand
8. Back Cover Plate 9126000-102
9. Electric Panel 9126000-103
10. Decorative Cap 8123000-202
Before attaching the lamp bulb, remove the parts that may drop
such as the filter and specimen from the microscope frame, and
place the microscope in its back so that the bottom plate is
exposed.
1. Pull the lock knob (1) on the bottom to opera lamp housing door
(fig.15).
2. Hold the halogen bulb (2) without taking it out of the polythyle-
ne bag so as not to taint the bulb with fingerprints and puch the
bulb into the pin holes on the socket (3). After attaching,remove
the polyethylene bag.
3. With the lock knob pulled out, close the lamp housing door,
then push the lock knob back to lock the cover.
Always use the designated bulb. Using a bulb other than those
specified by LABOMED may lead to a fore hazard or and low
illumination. Fingerprints or stains on the lamp bulb reduce its
life. If contamination occurs, wipe bulb surface with a cloth
slightly moistened with alcohol.
Figure 16
1. Pull out two decorative caps (shown as 9) from back cover
plate.
2. Using Cross head screw driver, unscrew two screws (shown as
9) to detach back cover plate from stand.
3. Lay down the microscope on back cover over a clean surface to
expose base plate.
4. Using Cross head screw driver, unscrew four screws (shown as
1) to open the bottom plate of the microscope.
5. Cut LED wire holding tie and Detach LED wire connector from
PCM mounted on back cover plate.
6. Detach Heat Shield from critical mount by opening 3 screws
(shown as 3) with cross head screw driver.
7. Detach LED from Heat Shield by loosening 2 screws (shown as
6).
8. Replace new LED on Heat Shield and connect wire connections
to PCB.
9. Reverse steps 6 to 1, to complete the process.
CAUTION: For Bulb Replacement During Use or Right After Use
The bulb, lamp socket and areas near these will extremely hot during and right after use. Set the main
switch to “O” (OFF), disconnect the power cord from the wall outlet, and allow the bulb ans lamp socket
to cool before replacing the bulb with a new bulb of the designated type. Cooling time may vary to users
Discretion.
9126000-795
Lx 400
Issue 1.5