SL 20i
GUIDANCE TABLES (continued)
23
Table 206– Recommended Separation Distances between
ME Equipment and ME Systems that are NOT Life-supporting.
Portable and Mobile RF Communications Equipment and the SL 20i
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
Recommended Separation Distances for between
Portable and Mobile RF Communications Equipment and the SL 20i
Max Output Power
of Transmitter
(W)
Separation (m)
150kHz to 80 MHz
d=(3.5/v1)(Sqrt P)
Separation (m)
80 to 800 MHz
d=(3.5/E1)(Sqrt P)
Separation (m)
800MHz to 2.5GHz
d=(7/E1)(Sqrt P)
0.01
0.1166
0.1166
0.2333
0.1
0.3689
0.3689
0.7378
1
1.1666
1.1666
2.3333
10
3.6893
3.6893
7.3786
100
11.6666
11.6666
23.3333
8126700-795
SL 20i
Issue 1.0
Printed on June,2019
The SL 20i is intended for use in the electromagnetic environment in which radiated RF
disturbance are controlled. The customer or user of the SL 20i can help prevent Electromagnetic
interference by maintaining a minimum distance between portable and mobile RF Communications
Equipment (transmitters) and the SL 20i as recommended below, according to the maximum output
power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note 1: At 80MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.