CAUTION! – Hazardous Gases!
If
the cabinet is connected to any external gas supply then this must be shut off prior to
any
service or repairs.
Ensure the work area is adequately ventilated
Read any data sheets for gas supply / fittings
Report any faulty or suspect supply pipes to cabinet user / facility manager.
NOTE
Recycling
All cabinet components except the HEPA filters can be recycled.
If the cabinet has been used to process hazardous biological or chemical material (powders
etc) then it must be fully decontaminated prior to disposal and a Certificate of
Decontamination must be produced by the operator before removal from the laboratory.
NOTE
– Start up after service / repair
If any safety devices (shielding / earthing etc) were removed or disabled prior to or during repairs then
the cabinet must not be started up until these devices are re-installed and checked for correct
operation. Before the cabinet is operated fully, it is recommended that a test run be carried out.
1.2 Standards and safety regulations.
The cabinet complies with, and has been tested to, the following standards and directives:
EN 12469:2000,
DIN 12980/2005/6
EN 292-1, EN 292-1,
EN 50081-1, EN 50082-1, EN 50081-2, EN 50082-2, EN 55011
IEC 801-2, IEC 801-4, IEC 1010-1 / EN 6101-1, UL 3101-1
EMC Directive 89/336 EEC-93/68/EEC
Low Voltage Directive 73/23 EEC-93/68/EEC
ENV 50140, ENV 50141
NFX 44-201