Mars-Mars PRO
User Manual
Code No. 9.900.900.710
U:\QA\Manualer\Arkiv\Manuals- Labogene-Origio-GKO\Manualer-WORD\ScanLaf\Mars Mars Pro manual-UK - rev. 06.doc
Page 31 af 34
The fluorescent and UV lamp (germicidal) have an average life time of about 5000 hours. SCANLAF,
however, suggests its replacement after no more than 2000-3000 hours. The luminous potential
delivered drops to 70% after said period.
REPLACEMENT OF THE FLUORESCENT LIGHT TUBES
Remove the air distributor located in top of the work chamber (handle with care)
Release the two lamps from the fixtures by rotating them 90º.
Install the new lamps performing the inverse operation.
REPLACEMENT OF THE UV lamp TUBES
The UV light is located on the back wall
Release lamps from the fixtures by rotating it 90º.
Install the new UV -lamp performing the inverse operation.
NOTE: Use gloves when manipulating a germicidal UV lamp, to prevent leaving hand prints that
reduce the effects of the ultraviolet radiation.
Replace UV lamp every 2000 or 3000 hours of life.
12.5 Spare parts
Cat. No.
Item
9.000.040.001
UV-light tube for Mars/Fortuna 900-1800
9.000.040.011
Light tube for Mars/Fortuna 900
9.000.040.012
Light tube for Mars/Fortuna 1200
9.000.040.013
Light tube for Mars/Fortuna 1500
9.000.040.014
Light tube for Mars/Fortuna 1800
9.000.050.001
Main HEPA filter for Mars/Fortuna 900
9.000.050.002
Main HEPA filter for Mars/Fortuna 1200
9.000.050.003
Main HEPA filter Mars/Fortuna 1500
9.000.050.004
Main HEPA filter Mars/Fortuna 1800
9.000.050.005
Main circuit board for Mars, Fortuna, Titan PCR
9.000.050.006
Flowsensor
9.000.050.007
Light ballast for Mars/Fortuna 900 1200