Instruction manual for CoolSafe 

 

 

01012016QA01 

Item No. 700637 

CopyRight © 2016

 

Page 

9

 of 

30

 

 

A hose may be attached to the nozzle to minimize spillage. Remember to close the drain valve after 
use. 

4.2.6 

Pressure regulation valve (optional) 

The pressure regulation valve is placed on the front of the cabinet. It is a needle valve that is used 
for fine-tuning the flow of air or inert gas into the ice condenser. It is used to reduce the hysteresis 
on the pressure regulation. 

 

 

Never close a needle valve too hard 

– there is a risk of the needle welding to the valve seat due to 

the very tight fit. 

4.2.6.1  Using the pressure regulation 

The pressure regulation works from the display of the controller. Here a set point of the pressure 
during the freeze drying is defined. 
 
If a set point of e.g. 1 hPa is entered, the controller will try to maintain the pressure of 1 hPa during 
the freeze drying process. 
 
So when the vacuum valve is open and the vacuum pump reduce the pressure to 1 hPa, then the 
pressure regulation valve will open, in order to inject air or inert gas to the freeze dryer chamber. 
However,  when  the  regulation  valve  open,  then  the  pressure  will  increase  very  fast  and  with  no 
control. To enable a controlled pressure increase, a needle valve is built in, enabling fine tuning of 
the flow into the chamber. 
 
This will prevent an over-shoot in the chamber pressure, and enabling the control of the pressure 
during the process. 
 
In order to get the best result of the pressure regulation and the needle valve, it is necessary to make 
a fine tuning of the needle valve. Same applies in case the set point pressure is changed. 
 
So whenever the set point is changed, the needle valve must be fine-tuned. 
 
If the needle valve is closed too much, then the pressure will still decrease, in which case the needle 
valve must be opened more, or if the pressure increases too fast, the needle valve must be closed 
more. 

4.3 

Interface, data acquisition and control  

On the back of the CoolSafe cabinet, to the right of the name plate, slot for a SD card and a mini 
USB connection is placed.  
 
During freeze drying all time, pressure and temperature data is stored on the SD card at one minute 
intervals. The data saved on the SD card is formatted for use in the LaboGene freeze drying software 
package. From the software the data can be formatted and saved in an Excel compatible format. 
The software package is included in the delivery. 
 
Under the SD card slot a mini USB connection is available. This is used for connecting a computer, 
running the LaboGene freeze drying software, from which the freeze dryer can be controlled. The 
software offers a graphical user interface showing the complete freeze dryer system and the status 

Summary of Contents for CoolSafe 100-9 Touch Superior XS

Page 1: ...Item No 700637 CopyRight 2016 INSTRUCTION MANUAL CoolSafe 100 9 Touch Superior XS CoolSafe 95 15 Touch Superior XL...

Page 2: ...dium or minor injuries or physical damage This document is the original instruction manual in accordance with the current Machinery Directive This document is the reference for any translation of this...

Page 3: ...essure 13 6 3 Auto menu 13 6 3 1 Programme list 14 6 3 2 Start 14 6 3 2 1 Current shelf temperature 14 6 3 2 2 Alarm pressure 15 6 3 3 Edit programme 15 6 3 3 1 Pre freeze 15 6 3 3 2 Alarm pressure 16...

Page 4: ...n Please read this entire manual to fully understand the safe and effective use of this product Should you have any comments about this manual we will be pleased to receive them Note If this equipment...

Page 5: ...um pump with 2 5 m3 h capacity or greater must be connected to the CoolSafe via the connection on the rear of the CoolSafe cabinet in order to allow cryo pumping The vacuum pump can be installed on th...

Page 6: ...n easily be controlled via the control panel or user interface 4 1 Control panel The control panel is placed on the upper front of the cabinet at an angle for easy reading and use The control panel a...

Page 7: ...e freeze dryer function and operate correctly Each must be set correctly and plugged in correctly to make the CoolSafe operational 4 2 1 Power supply and ON OFF switch The CoolSafe is designed to be p...

Page 8: ...ion purposes the temperature in each shelf is measured using a PT 100 temperature sensor The signals from these sensors regulate the heating wire build into each shelf 4 2 4 Product temperature The te...

Page 9: ...ilt in enabling fine tuning of the flow into the chamber This will prevent an over shoot in the chamber pressure and enabling the control of the pressure during the process In order to get the best re...

Page 10: ...to facilitate easy control of the freeze dryer and the freeze drying process in either manual or automatic running mode The touch control system is based on a simple structured menu structure for easy...

Page 11: ...Action When entering the manual menu the compressors start cooling the ice condenser Scroll up and down using the arrows to the right to select the menu item you want to see or change Freeze drying t...

Page 12: ...press to reset error counter Press the LaboGene logo to return to the main menu Press the LaboGene log for 2 3 seconds to lock the display for input Unlock press again 6 2 1 Current shelf temperatures...

Page 13: ...ot turned on before this pressure is reached If the pressure raise above this pressure during operation the shelf heating will be turned off to protect the product being freeze dryed Decrease set poin...

Page 14: ...ure Actual pressure View the current pressure in the ice condenser Actual shelf temp Press to see list of Current shelf temperatures Product temp View the current product temperature Set point shelf t...

Page 15: ...Pre freeze Press to edit pre freeze process Alarm pressure Press to set alarm pressure Primary drying 1 5 Press to edit primary drying process Primary drying 6 9 Press to edit primary drying process S...

Page 16: ...3 seconds to lock the display for input Unlock press again 6 3 3 2 Alarm pressure Touch screen as opened Item Action Decrease set point by 0 1 hPa Torr Accept change in set point Increase set point by...

Page 17: ...Gene logo to return to the primary drying menu Press the LaboGene log for 2 3 seconds to lock the display for input Unlock press again 6 3 3 4 Secondary drying Touch screen as opened Item Action When...

Page 18: ...process Heating elements in the condenser will start melting the ice on the condenser wall Can either melt it completely or just enough to free the ice block from the condenser wall The block can the...

Page 19: ...d the release valve will be closed 6 7 Setup Touch screen as opened Item Action Language selection Press to select the language displayed in the control system Pressure selection Press to select the p...

Page 20: ...LaboGene logo to return to the setup menu Press the LaboGene log for 2 3 seconds to lock the display for input Unlock press again 6 7 4 Set time date Touch screen as opened Item Action When activatin...

Page 21: ...u Press the LaboGene log for 2 3 seconds to lock the display for input Unlock press again 6 9 Info Touch screen as opened Item Action Software version controller The current software version installed...

Page 22: ...minutes At start up stage 1 is running for 6 minutes before stage 2 is started up This is to ensure that pressure in the stage 2 system is keep within the allowable pressure limits In case the ice con...

Page 23: ...te period it must be reset by powering the compressor down and turning it on again 8 Preparations Check that the CoolSafe ice condenser is free from ice and water and that all valves i e the drain val...

Page 24: ...reezing time and shelf temperature 5 steps If you do not need 5 steps enter 0 for the time in the steps not used In the pre freeze mode vacuum is not applied Enter the values for freeze drying time sh...

Page 25: ...arly during the week it is recommended that the vacuum pump run constantly with open gas ballast valve and closed vacuum valve 8 4 Hints tips and reminders The selection of the appropriate sample cont...

Page 26: ...he condenser has dried 9 4 Checking for Leaks on the CoolSafe freezer dryer Always start from the vacuum pump and work forwards towards the chamber 1 Check vacuum hose connection to the pump If the pu...

Page 27: ...inciple the process is very simple but in reality it also can be very complex as the product can consist of many components which interact with each other Moving from one state to another requires the...

Page 28: ...s kg 4 Total volume of Condenser L 9 Weight net kg 75 Cabinet dimensions DxWxH mm 620x495x745 Cabinet material Polyester coated steel Condenser material Stainless steel AISI 316 Cooling media R507 R17...

Page 29: ...Instruction manual for CoolSafe 01012016QA01 Item No 700637 CopyRight 2016 Page 29 of 30 13 Declaration of conformity...

Page 30: ...46 13 0 1 986 22 2 0 835 44 0 0 081173 4 0 4 371 13 2 1 951 22 4 0 819 46 0 0 064112 4 2 4 298 13 4 1 915 22 6 0 803 48 0 0 050383 4 4 4 226 13 6 1 880 22 8 0 789 50 0 0 039387 4 6 4 154 13 8 1 847 23...

Reviews: